![TSI Incorporated aerotrak 9306-04 Скачать руководство пользователя страница 55](http://html1.mh-extra.com/html/tsi-incorporated/aerotrak-9306-04/aerotrak-9306-04_operation-manual_1178070055.webp)
Operation
3-31
Field
Description
Room Area
Displays the area of the room in ft
2
or m
2
.
Class Level
Depends on the report definition, see below.
Air Flow
Displays the airflow characteristics of the room.
Room Status
Displays the status of the room. See
Room Definition
Screen
below.
Min Locations
Displays the minimum number of locations that must be
sampled in the room.
Min Samples
Displays the minimum number of samples that must be
taken at each location.
Min Vol. per channel
Allowable channel sizes for the selected Class Level for
that Standard.
Room Definition
Press to set definitions for the room. (See
Room
Definition Screen
below.)
Generate
Press to begin generating a report that you can view on-
screen or print. (See
Generate Screens
below.)
Room Definition Screen
Use this screen to define specific values for the room. Press
OK
when finished.
Field
Description
Room Status
Select the room status: As Built, At Rest, or
Operational.
Air Flow
Select the air flow: Unidirectional or Non-unidirectional.
Class
Select the class of the room: The class is dependent on
the standard:
FED FT3: 1, 20, 100,1000,10000, 100000
FED M3: M1.0, M1.5, M2.0, M2.5, M3.0, M3.5, M4.0,
M4.5, M5.0, M5.5, M6.0, M6.5, M7.0
ISO14644-1: 1, 2, 3, 4, 5, 6, 7, 8, 9
EC GMP: A, B, C, D
Area
Use the on-screen keyboard to enter the area of the
room in ft
2
or m
2
.
netzerotools.com
netzerotools.com
Содержание aerotrak 9306-04
Страница 2: ...ii netzerotools com netzerotools com ...
Страница 6: ...netzerotools com netzerotools com ...
Страница 54: ...3 30 AEROTRAK Handheld Airborne Particle Counter netzerotools com netzerotools com ...
Страница 57: ...Operation 3 33 netzerotools com netzerotools com ...
Страница 70: ...7 2 netzerotools com netzerotools com ...
Страница 81: ...netzerotools com netzerotools com ...