Troyka Med Inc 

IOSB Mah. 2284. Cad No: 48 Yenimahalle Ankara TURKEY  

Tel:

 +90 5325128580- [email protected] – 

www.troykamed.com

 

T-280 MRI Conditional Dom Camera Op. Man. Pub. 022018 Pub. No: 0 Rev:0  

 

Warning! 

Self-heating of cables due to improper cable routing! Injury to equipment and/or 
operating personnel. Cables should not be looped crossed inside MRI room. 

 
 
 

 

1.4 Warnings for MRI  

 

Caution! 

Installation of materials inside the MRI must be done with extreme caution. 

 

Caution! 

Take care that ferromagnetic materials be kept at least three meters away from the 
magnet and that no installation shall be done near the filter panel if a scan is in 
progress.  

 

Caution! 

In addition, no persons with ferromagnetic prosthetic devices, such as pacemakers or 
joints replacement, should enter the MRI suite at any time. Extreme, high magnetic 
fields inside the magnetic room have the potential to dislodge items at high velocities 
and can result in serious injury, or death.   

 

Caution! 

For questions regarding installation procedures or technical support, call Troyka Med 
Service team or contact via email at 

[email protected].

 

 

 

Caution! 

Only system components explicitly designed for use inside the MRI suite should be 
placed inside the magnet room. Components not designed for MRI use may present a 
projectile hazard and can become airborne, causing serious injury, damage or death.  

 

2 General Information  

2.1. Inspection of Delivered Goods 

Each of MRI dom camera has been thoroughly tested prior to deliver and is ready for immediate use. Upon 
receipt, please report any transportation damage or missing accessories immediately. Troyka Med can only 
accept liability for such damages if it is claimed prior to initial operation. In case of transportation damage, 
please contact the Troyka Med service department. For faster support please have ready all shipment details 

Содержание T-280 MRI

Страница 1: ... No 48 Yenimahalle Ankara TURKEY Tel 90 5325128580 info troykamed com www troykamed com T 280 MRI Conditional Dom Camera Op Man Pub 022018 Pub No 0 Rev 0 OPERATORS MANUAL T 280 MRI Conditional Dom Camera MRI CONDITIONAL TO 3 T 0 TESLA ...

Страница 2: ...485 2016 To ensure that you obtain maximum benefit from the T 280 MRI Dom Camera please take a few minutes to read the enclosed information regarding operation service and maintenance After reading this manual store it in a safe place for future reference If you have any problems in the meantime or would like any advice about this or any other MR products from the Troyka Med range please contact u...

Страница 3: ...table risks to the patient medical staff or other persons within the MR environment 1 2 General Safety Information and Intended Use The MRI dom camera must only be operated by personnel property trained in MRI safety The MRI dom camera must only be operated by personnel properly trained in identifying interference problems such as artifacts streaks and distortions in image data It is required that...

Страница 4: ...the dom camera possible electric shock or fire Don t handle the power cable with wet hands High voltages present could cause lethal electric shock Warning Should any foreign substances enter the device such as liquid metal chips or dust immediately turn off power to the Interface Unit Under no circumstances shall tools or foreign objects be inserted into the device as this could result in failure ...

Страница 5: ...ields inside the magnetic room have the potential to dislodge items at high velocities and can result in serious injury or death Caution For questions regarding installation procedures or technical support call Troyka Med Service team or contact via email at info troykamed com Caution Only system components explicitly designed for use inside the MRI suite should be placed inside the magnet room Co...

Страница 6: ...ase consult your Troyka Med Sales or Service office Always use the power supply cord supplied in the original shipping carton If the enclosed power cord could not be used due to a different standard in your country use a power cord that conforms to the fallowing regional standards United States UL United Kingdom BASEC BS Germany VDE Switzerland SEV Canada CSA Japan MITI In other regions please use...

Страница 7: ...or overview MRI room The T 280 MRI Dom Camera System is designed to be MRI conditional and is manufactured and tested to the highest standards The camera system is designed to be used with all MRI scanners up to 3 Tesla magnetic field strength A picture of MRI conditional dom camera is given in Figure 1 The camera can be used both on the wall or ceiling mount position and the angle of view could b...

Страница 8: ...nic Parameter 1 25s 1 100 000s 25fps 1080p 30fps 1080P High speed long distance real time transmission Full HD and standard definition switch Day night ICR 3D DNR 3 4 Installation 3 4 1 Installation Diagram of MRI Video Camera in Magnet Room A representative drawing is showing the layout plan of the MRI video camera is given in Figure 2 The camera shown in Figure 2 1 depending on the customer requ...

Страница 9: ...em will be ready to use In order to power on the camera turn on the power switch which is available on the interface unit Figure 2 9 Figure 2 MR conditional dome type video camera installation diagram 3 4 2 MRI Camera Installation diagram The representative drawing showing the installation diagram of the MRI camera is given in Figure 3 Use 4 screws delivered in the package to mount the base part o...

Страница 10: ...g Unplug all power connections before cleaning 4 1 1 Monitor Surface Dampen do not saturate a clean lint free cloth with environmentally friendly glass cleaner The cleaner contains spirit as an active substance up to 98 and tensides which can be biologically recycled Cline the monitor screen using circular motions with the cloth to avoid streams Remove fingerprints grease dirt and dust Carefully d...

Отзывы: