Troxler TroxAlert Скачать руководство пользователя страница 3

 

TroxAlert

 Radiation Survey Meter 

3

 

INTRODUCTION 

 
The TroxAlert

 Radiation Survey Meter provides an 

inexpensive means of fulfilling the survey requirements of an 
ALARA (

A

L

ow 

A

R

easonably 

A

chievable) program. The 

TroxAlert is a basic tool that can be used to survey storage areas 
and work sites to ensure a safe working environment. 
 
The TroxAlert uses a Geiger-Mueller (G-M) tube to detect 
alpha, beta, gamma, and x-ray radiation. The measured radiation 
is shown on the scale and on the LED on the front of the case. 
The TroxAlert also beeps when radiation is detected.  
 

NOTE 

TroxAlert reactions to natural background 
radiation are normal and expected. 

 
The TroxAlert has two switches: the 

ON–OFF–BATT

 rocker 

switch on the left-hand side of the case and the 3-position range 
switch on the front of the case. 
 
The range switch is used to set the measurement range of the 
TroxAlert. With the switch in the 

X1

 position, the scale 

measures from 0 to 1 mrem per hour (0-10 

µ

Sv/h). In the 

X10

 

position, multiply all readings by 

10

 [0 to 10 mrem/h 

(0-100 

µ

Sv/h)]. In the 

X100

 position, multiply all readings by 

100

 [0 to 100 mrem/h (0-1000 

µ

Sv/h)]. 

Содержание TroxAlert

Страница 1: ...diation Survey Meter Troxler Electronic Laboratories Inc and subsidiary Troxler International Ltd 3008 Cornwallis Road P O Box 12057 Research Triangle Park North Carolina 27709 Tel 919 549 8661 Fax 91...

Страница 2: ...his document may be reproduced or transmitted in any form or by any means electronic or mechanical including photocopying recording or information storage and retrieval systems for any purpose without...

Страница 3: ...on the scale and on the LED on the front of the case The TroxAlert also beeps when radiation is detected NOTE TroxAlert reactions to natural background radiation are normal and expected The TroxAlert...

Страница 4: ...bove turn the meter power on and set the range switch to the X10 position change the range as needed Hold the meter about 1 foot 0 3 m away from the gauge and point the meter at the gauge Slowly bring...

Страница 5: ...r detects radiation or until the switch reaches the X1 position CAUTION Do not allow the TroxAlert to come in contact with a radioactive source Such contact may contaminate the TroxAlert and render it...

Страница 6: ...ional Institute of Standards and Technology NIST is used for this calibration The accuracy of any single reading based on this calibration is 20 Note that environmental conditions and user technique a...

Страница 7: ...1 5 2 0 mg cm2 Detectable Alpha beta gamma and x ray Radiation Display Range 0 1 mrem h 0 10 Sv h 0 10 mrem h 0 100 Sv h 0 100 mrem h 0 1000 Sv h Accuracy 20 of indicated reading Cs 137 gamma Calibra...

Страница 8: ...Battery Life Up to 2000 hr normal operational conditions MECHANICAL SPECIFICATIONS Case Size 5 75 L x 3 25 W x 1 5 D in 14 6 L x 8 26 W x 3 8 D cm Weight 7 68 oz 0 22 kg with battery 6 08 oz 0 17 kg w...

Отзывы: