STEP 5 – TAKE MEASUREMENTS 
Always begin with handle in SAFE position

. Ensure 

that the handle is in the SAFE position by firmly 

tapping down on the handle without engaging the 

trigger. 
We recommend 2 minutes for measurement time and 

1 minute for background time. (Setup menu) 
The gauge can be set to receive moisture values from 

an outside device or by keypad entry.  You can also 

choose to perform the moisture measurement before 

or after the density measurement. Press 

Setup

,

 

choose 

4. 

Moisture Input

, then choose your method / order 

desired. 

Prepare the test site

 as described in step 4. If using 

the moisture probe, drive the drill rod to at least 15 cm 

(6 in).  

PERFORMING DENSITY MEASUREMENTS 

1

1

.

.

 

 

Place the gauge on the test site with the 

handle in the Safe position and press 

ENTER/START

. Follow the prompts on 

the screen to either lower the handle to the 

desired measure depth or wait for the 15-

second initialization phase to complete (this 

will occur approx. every 10 min. or as 

programmed). Then lower handle to the 

desired measurement depth, carefully 

seating the handle in the proper notch and 

press 

ENTER/START

 

2

2

.

.

 

 

After count time, (when prompted) select 

the active background count by answering 

Yes

 to the displayed question or take a new 

one by answering 

No

 to the question.  If 

taking a new background count, move the 

source rod to background position and 

press 

ENTER/START

. For accurate density 

measurement of soils and aggregates, a 

new background count should be taken at 

the first test site of the day and on a new 

material or a new jobsite during the day.  

3

3

.

.

 

 

After the measurement, raise the handle.  

The gauge displays the measurement 

results or prompts for the moisture reading.

 

If a moisture reading is to be done, see the 

next section. To store the reading, press 

STORE

After taking readings, lift the gauge from the test site 

by the source rod handle. This returns the source rod 

to the 

SAFE

 position. When not taking readings, 

always keep the source rod in the 

SAFE

 position.  

PERFORMING MOISTURE MEASUREMENTS IN 

SOIL MODE 

1

1

.

.

 

 

When prompted by the 4540 gauge, insert 

moisture probe into the hole until the base 

plate contacts soil surface. The probe should 

fit tightly in the hole, some resistance should 

be felt when lowering it. 

2

2

.

.

 

 

Power the probe on (green light illuminates). 

The blue light will remain solid when Bluetooth 

connectivity is established.  

3

3

.

.

 

 

Press 

Enter

 on gauge keypad to perform 

moisture reading, wait a few seconds then 

press 

Enter

 on the keypad again to accept 

the moisture reading. See the Guide for the 

Model 6760 Moisture Probe for more 

details on operation. 

4

4

.

.

 

 

Carefully remove the probe and set aside. 

USING THE OFFSET FUNCTION 

The gauge readings can be adjusted using an offset. 

The gauge applies the offset to measurements until 

the offset is disabled or the gauge is turned off.  The 

gauge provides moisture 

density 

and 

trench

 offsets. 

Press 

OFFSET

 to access the 

Offset

 menu.  

MORE INFORMATION 

The user manual provides more information about the 

gauge’s functions and use. 

 

i. Model 6760 Quick Reference Guide 

ii. 6760 Technical Notes 

 

Model 4540 Quick Reference Guide PN 128992, Edition, 1.1. May 2021 

 

Model 4540 

Soil & Asphalt Density Gauge  

 

E

Gauge

 Combo 

QUICK 

REFERENCE 

GUIDE 

 

 

 

 

 

 