Open the valve and apply test gas at a rate of
0.5 litre/min.
Wait for the reading to settle at the stable value
(Not necessarily the test gas value).
Press
L
and
Span
the displayed value. If the
displayed value continues to shift press
L
again.
Shut the valve and stop the supply of test gas.
Disconnect the application tube from the gas
hood and remove the gas hood.
4.5.1.3 Calibrate Span - Using Test Gas of a
Different Value.
The screen will show the concentration of
Span Gas
that is expected to be used.
Check the value of
Span Gas
displayed and
compare it against the value shown on the test
gas cylinder. If the values do not match press
L
.
Press
L
to increment each digit
Press
R
to move to the next digit.
When the last digit is set, press
R
to select
Done
and return to the
Calibrate Menu
.
Continue with
Calibrate Span
as described in
the previous section
Calibrate Span - Using Test
Gas of Current Value
.
Содержание Sentro Wireless TX6355
Страница 1: ...TX6355 Sentro Wireless...
Страница 2: ......
Страница 6: ...6 P5596 1600 rev B www trolex com 1 4 Dimensions...
Страница 10: ...10 P5596 1600 rev B www trolex com...
Страница 11: ...P5596 1600 rev B 11 www trolex com TX6355 User Manual...
Страница 18: ...18 P5596 1600 rev B www trolex com 4 Setup and Calibration 4 1 Controls and Indicators...
Страница 19: ...P5596 1600 rev B 19 www trolex com TX6355 User Manual...
Страница 20: ...20 P5596 1600 rev B www trolex com 4 2 Software Menus...
Страница 56: ...56 P5596 1600 rev B www trolex com...
Страница 57: ...P5596 1600 rev B 57 www trolex com TX6355 User Manual...