7. TECHNICAL INFORMATION
7.1 Product specification
Sensing technology
Optical Refraction Technology (ORT)
Light-Scatter Photometer (OPC)
Particulate measurement
Target RCS identification range 1µm – 10µm
Max. typical dust loading*
150mg/m³
Continuous range
25mg/m³
Displayed data
RCS mg/m³
Total particles/litre
Resolution
1,000th of a mg
Averaging period
15 minutes, 1 hour, 4 hours, 8 hours and 12 hours (rolling avg.)
Sampling interval
1
– 60 seconds
Particle count
>600 particles/second
Total airflow rate
~1.5L/m (nominal)
Typical RCS accuracy
+/- 25%
*The instrument can define particulate measurement peak trends up to the quantity specified.
Note:
Sustained exposure to dust quantities above 25 mg/m³ will be logged; however, it may affect
the operating life of the Air XS sensor.
Particulate sensing parameters