 Revised by V. Li 01/25/2019 

(although a minimum gauge of 16 AWG has been acceptable).  This bonding 

connection is necessary to prevent internal damage to the power module in the 

event of any accidental reversed polarity connections or accidental connections 

between the battery positive wiring and the case of the power module. 

8. A Remote User Interface (RUI) module must be connected to the power module

using two CAT-5 cables.  The COMM1 jacks should be connected together and

the COMM2 jacks should be connected together.  If the cables are not connected

or if they are interchanged, the power module will not operate.  The RUI may be

either a Tripp Lite RUI or a customer designed RUI that meets Tripp Lite

specifications.

9. The power module includes a low voltage cutoff (LVC) function that will remove

any load on the battery when the battery is depleted.  The LVC level is about

10.5V to assure optimum battery cycle life.  If any other DC loads are present in

the system, they need to include their own LVC function in order to prevent

damage to the battery due to over discharge.

10. The power module will not turn on unless connected to a battery that has a

voltage greater than 9.6V.  If the battery voltage is less than 9.6V, it is assumed

to be defective and the charger is inhibited in order to prevent possible

outgassing of the battery.

11. The power module reports % Battery Capacity through its USB port.  For

HC350S, this data is most accurate when a 55Ah sealed lead acid absorbed

glass mat (AGM) battery is used.  For HC350S1, this data is most accurate when

two 33AH sealed lead acid absorbed glass mat (AGM) batteries are used.

Batteries of other capacities may be used, but the accuracy of the % Battery

Capacity report will be reduced.

12. After connecting a battery to the power module, the power module should be

allowed to charge the battery to full capacity in order to initialize the % Battery

Capacity report and the % Battery Capacity LED’s (on the RUI). This will take at

least 4 hours, even with fresh battery.  The % Battery Capacity report and LEDs

will not be accurate until after the initial charge has been completed.  After

initialization, the power module will properly indicate the % Battery Capacity as

the battery is repeatedly discharged and recharged.  However, each time the

battery is replaced, the initialization will need to be repeated.

13. For IEC 60601-1 4

th

 edition compliance, external clamps as circled (shipped

loose with the power module) need to be installed to the DC input and the AC

output as shown: