![Tripp Lite AG-8862 Скачать руководство пользователя страница 2](http://html1.mh-extra.com/html/tripp-lite/ag-8862/ag-8862_owners-manual_1167491002.webp)
2
Important Safety Instructions
SAVE THESE INSTRUCTIONS
This manual contains instructions and warnings that should be followed during the installation,
operation and storage of all Tripp Lite UPS Systems. Failure to heed these warnings may affect
your warranty.
Statement of Intended Use
Tripp Lite Medical-Grade UPS Systems are intended to support and protect non-medical and
medical computer equipment and medical devices that require leakage current reduction, surge
protection, voltage regulation, line noise filtering and battery backup during power outages and
generator testing, both inside and outside patient care areas. Tripp Lite Medical-Grade UPS Systems
come with hospital-grade plugs and receptacles that reduce leakage to below 100μA.
Product Information
SMART700HGL
SMART1200XLHGL
AC Input Voltage
120
120
AC Input Voltage Range
81–143
81–143
Input Amp(s)
5.5
9.2
Output Amp
5.8
6.25
Rated Power (VA/Watts)
700VA/450W
1000VA/750W
Frequency
60 Hz
60 Hz
No. of Phases
Single
Single
Class Type
Class I
Class I
Plug/Connector Type
C14 Inlet Dongle, 6 in.
C14 Inlet Dongle, 6 in.
Operation
Continuous
Continuous
AC Input Protection (Qty./type/rating)
Input breaker
(2 x 10A), thermal
Input breaker
(2 x 15A), thermal
Max Leakage Current (uA)
Less than 100uA
Less than 100uA
Weight
33.7 lb. (15.3 kg)
33.6 lb. (15.2 kg)
Known Contraindication
None
None
UPS Location Warnings
Relevant Symbols
Denotes that information in the manual should be reviewed before use.
Used to signal a warning that a statement is particularly important and could pose a safety
risk.
Warning—Dangerous Voltage
No sitting
Denotes general warning sign
No stepping on surface
Refer to Instruction Manual/Booklet
No pushing
18-06-068-933608.indb 2