Earth Voltage Test 

 

1.  Set the Function selector switch to the EARTH VOLTAGE position (AC or DC) 

 

2.  Connect the 2 Voltage test leads to the meter.Connect the black and red leads - connect the 

black lead to the E (green jack) and connect the Red test lead to the C (Red jack). 

 

3.  Connect the test leads to the item under test. 

 

4.  Confirm that the voltage measurement is less than 10V AC; otherwise accurate Earth 

resistance measurements cannot be made. If voltage is present (higher than 10V AC), the 

source of the voltage must be found and corrected before testing can continue. 

 

Earth Ground Resistance test 

 

•  Connect the 3 test leads to the meter (1) as follows: 

Green lead (2) to the 'E' terminal 

 

Yellow lead (3) to the 'P' terminal   
Red lead (4) to the 'C' terminal 

•  Drive the Auxiliary Earth Bars P1 (6) & C1 (7) into the ground. Align the bars an equal distance 

apart to the existing Earth ground rod connection and in a straight line as shown in the diagram 

above. 

 

If the auxiliary bars are placed in close proximity to the ground stake, 

measurement inaccuracies will result. 

(Minimum distance between rods should be no less than 10 feet (3m). 

 

•  Connect the alligator clip end of the test leads to the Earth bars and existing ground rod 

connection as shown above: 

 

Green lead (2) to the existing ground rod 
(5) Yellow lead (3) to Earth Bar P1 (6) 

Red lead (4) to the Earth Bar C1 (7) 

 

•  Set the meter's function switch to the desired resistance range. (20, 200, 2000Ω). 

 

•  Zero the range as described in the Zero adjustment procedure above. 

 

•  Press the TEST button. . The “   “ icon will flash and the audible signal will sound. 

 

•  Note the displayed reading. 

 

•  If high resistance is detected, note the value and take appropriate steps to correct the 

ground connection if necessary. 

 

•  Press the TEST key to end the test. 

 

•  Readings of “1” are typical when the test leads are not connected to the meter. 

 

www.calcert.com

[email protected]

1.800.544.2843

0

5

10

15

20

25

30

Содержание EG480

Страница 1: ...fully before attempting to operate the meter and use the meter only as specified in this manual Environmental safety information Do not use meter outdoors when precipitation is likely Installation Category III Pollution degree 2 Altitude 2000 meters max Ambient conditions 32 to 104oF 0 to 40oC RH 80 max Observe the following international safety warning symbols Safety Symbols Caution Refer to this...

Страница 2: ...Function Selector switch Selects desired range and function Top View 1 V C Input Red 2 P Input Yellow 3 COM E Input Green Operation ZERO Adjustment of Earth Resistance Ranges 1 Connect the long test leads to the meter as follows Green lead to the E terminal Green jack Yellow lead to the P terminal Yellow jack Red lead to the C terminal Red jack 2 Set the function selector switch to the desired mea...

Страница 3: ...ry Earth Bars P1 6 C1 7 into the ground Align the bars an equal distance apart to the existing Earth ground rod connection and in a straight line as shown in the diagram above If the auxiliary bars are placed in close proximity to the ground stake measurement inaccuracies will result Minimum distance between rods should be no less than 10 feet 3m Connect the alligator clip end of the test leads to...

Страница 4: ... 750V AC position 2 Connect the test probes to the circuit under test 3 Note the displayed Voltage value DC Voltage Measurement 1 Connect the red test lead to the V C connector and the black test lead to the COM E connector 2 Set the function selector switch to the 1000V DC position 3 Connect the test probes to the circuit under test 4 Note the displayed Voltage value 5 Three terminal Ground Resis...

Страница 5: ...he valid legal stipulations in respect of the disposal of the device at the end of its lifecycle Cleaning and Storage Periodically wipe the meter s case with a damp cloth and mild detergent do not use abrasives or solvents If the meter is not to be used for a period longer than 60 days remove the batteries and store them separately Specifications General Specifications Measurements Earth Ground Re...

Страница 6: ... to 500Hz 0 1V 3 reading 3digits Resistance 0 to 200kΩ 0 1k 1 reading 2 digits Overload Protection 250 Vrms AC Voltage 0 to 750V 1V 1 2 reading 10 digits 40 Hz to 400Hz Overload Protection 750 Vrms Input Impedance 10M 0 to 1000V 1V 0 8 reading 3 digits DC Voltage Overload Protection 1000 Vrms Input Impedance 10M www calcert com sales calcert com 1 800 544 2843 0 5 10 15 20 25 30 ...

Страница 7: ...for use that the products sold by it will be free from defects in workmanship and material for a period of 1 one year from the date of purchase This warranty does not apply to any of our products which have been repaired or altered by unauthorized persons in any way or purchased from unauthorized distributors so as in our sole judgment to injure their stability or reliability or which have been su...

Отзывы: