Page 3
symbOls,INDICaTORs,andDegReeOFsaFeTy
Attention: Consult accompanying documents.
Type B Equipment: Equipment providing a particular degree of protection against electric shock,
particularly regarding allowable leakage current and reliability of the protective earth connection.
This product is Class I equipment in which protection against electric shock does not rely on BASIC
INSULATION only, but which includes an additional safety precaution in that means are provided
for the connection of the EQUIPMENT to the protective earth conductor in the fixed wiring of the
insulation in such a way the ACCESSIBLE METAL PARTS cannot become LIVE in the event of a
failure of the BASIC INSULATION.
waRNINgsaNDUsagehINTs
Please make note of the meaning of the following warnings and usage hints:
Note: indicates usage information that helps the user to use the product correctly and efficiently or to
understand the properties of the product.
Protection against harmful ingress of water--Ordinary.
Not suitable for use in the presence of a flammable anesthetic mixture with air or oxygen or nitrous
oxide.
Maximum continuous operation of motor is 12 minutes per hour.
Input: 120 Volt AC, Single Phase, 60 Hz, 6.8 Amps.
WATER
ANESTHETICS
ELEC. RATING
OPERATION