trendmedic TM-8500 Premium Скачать руководство пользователя страница 7

 

Errors or Troubles 

 

     

Error 

message   

or Troubles 

   
                              Description 

 
        Remarks 

 

No display 

-Press and hold the TEST button until power on. 
-Check the battery   

   

   

 

Auto Off 

-Low battery 
-Over the period of test stage 

 

   

         
 
            Err 

-Blowing unsteady or interrupt 
-Blowing less than 6 seconds 
-Blowing flow too weak 
-No blow during BLOW mode. 
-Change temperature suddenly 

     

   

   

 

-Low battery indication 
-Need to change new battery 
-Use alkaline battery 

 

 

   
   

Alarm sound 

-During the blowing mode 
-Error message indication 
-Above alcohol set point   
-Above 500 tested times (Display counter when power on) 

 
   

   
   
    HI 
 

     
-The blood alcohol content above the display range. 
-Display range: 0.20%BAC(2.00‰, 1.000mg/L)

 

 

 

 

 

 

Important notes 

 

  Before testing, wait at least 15 minutes after your last drinking. 

  The blowing time must be enough until beeping stops.   

  It may show ERR (error) if none, interrupt or shorter blowing.   

  If the device changes temperature suddenly, wait some minutes to use.   

  Only use the device in a temperature 10°C ~ 40°C o r 50°F ~ 104°F.     

  Recalibration interval suggest: 12 months or 500 tested times 

  Never store the device in a corrosive gas room. 

  Never store the device under sun exposure area. 

  Avoid dropping or subjecting the device to a strong shock. 

 
 

Specifications 

 

 

Electrochemical sensor (Fuel Cell Sensor) 

 

Blowing time about 6 seconds 

 

Blowing flow about 10L/min 

 

Blowing flow interrupt detection 

 

Test counter to display the operated number when power on 

 

4 digits display with status and backlight 

 

Display range: 0.000-0.200%BAC(2.00‰ or 1.00mg/L) 

Содержание TM-8500 Premium

Страница 1: ......

Страница 2: ...4 Important Notes 5 Specifications 5 Calibration and Record 6 Precautions 6 Warranty 7 Introduction The device is intended to measure alcohol in the human breath Measurements obtained by this device a...

Страница 3: ...ng wait at least 15 minutes after your last drinking Only use the device in a temperature10 C 40 C or 50 F 104 F Description of parts Power Supply Slide back the battery cover on the back of the devic...

Страница 4: ...device may not use correctly You need to change a new battery Operation Test mode Insert one of the mouthpieces into the blow in hole Press and hold the TEST button until power on Release it after pow...

Страница 5: ...econds The device displays WAIT and flashing the 0 on the screen Your test result will be displayed on the screen for 20 seconds Your test result will be changed the color if above alarm point It will...

Страница 6: ...ice displays the reading for 20 seconds You press the memory button to check your last 5 tested records The device will be turn off after displaying 20 seconds The memory button is only used at this m...

Страница 7: ...Important notes Before testing wait at least 15 minutes after your last drinking The blowing time must be enough until beeping stops It may show ERR error if none interrupt or shorter blowing If the...

Страница 8: ...e unit of measure for mg L Power supply 2 pcs AAA 1 5V alkaline battery Battery for standard testing is about 500 times Low battery indication and auto power off function Operation 10 C 40 C or 50 F 1...

Страница 9: ...ght back to the local customer If a defect arises and we receives a valid claim within the Warranty Period at its option we will either 1 Repair the unit 2 Exchange the unit which is new or used funct...