Errors or Troubles
Error
message
or Troubles
Description
Remarks
No display
-Press and hold the TEST button until power on.
-Check the battery
Auto Off
-Low battery
-Over the period of test stage
Err
-Blowing unsteady or interrupt
-Blowing less than 6 seconds
-Blowing flow too weak
-No blow during BLOW mode.
-Change temperature suddenly
-Low battery indication
-Need to change new battery
-Use alkaline battery
Alarm sound
-During the blowing mode
-Error message indication
-Above alcohol set point
-Above 500 tested times (Display counter when power on)
HI
-The blood alcohol content above the display range.
-Display range: 0.20%BAC(2.00‰, 1.000mg/L)
Important notes
•
Before testing, wait at least 15 minutes after your last drinking.
•
The blowing time must be enough until beeping stops.
•
It may show ERR (error) if none, interrupt or shorter blowing.
•
If the device changes temperature suddenly, wait some minutes to use.
•
Only use the device in a temperature 10°C ~ 40°C o r 50°F ~ 104°F.
•
Recalibration interval suggest: 12 months or 500 tested times
•
Never store the device in a corrosive gas room.
•
Never store the device under sun exposure area.
•
Avoid dropping or subjecting the device to a strong shock.
Specifications
Electrochemical sensor (Fuel Cell Sensor)
Blowing time about 6 seconds
Blowing flow about 10L/min
Blowing flow interrupt detection
Test counter to display the operated number when power on
4 digits display with status and backlight
Display range: 0.000-0.200%BAC(2.00‰ or 1.00mg/L)