-16-
(8) Horizontal adjustment and fixed dental unit
①Screw six M12 hexagonal bolts into the M12 screw
thread holes in the bottom cover of the chair to the
level of chair balancing.
②Fixed the board by screws with the ground ,advoiding
the dental unit rollover during misoperation.
Note:
In the process of adjusting, make sure that bottom
cover touches the floor as much as possible to ensure
the stability of the whole set.
(9)instruments installation of tray
a. 3-way syringe
Take out the nozzle of the 3-way syringe, press the
meter ring and insert the nozzle, and then release the
meter ring to make it back to the original position.
If the meter ring cannot recoil to the origin, the
insert of the nozzle is not finished. The meter ring
should be pressed again to make the nozzle inserted
till the meter ring recoils.
b. hand piece
turn on the water source and air source before
installing the hand piece, pull out every hand piece
tubing and push on the footswitch, blow off the
remained water and air in the tubing, install the hand
piece to the tubing connector.
Note:
Please refer to the hand piece user manual as
hand piece using, disinfection and maintenance
(10)On side installation must be performed by the
qualified distributor authorized by ANYE.
(11)
Installation for detachable parts :
Fuse(see Page 11), water tank(
capacity 800ml, pressure 0.25MPa
),
strong suction,weak suction ,saliva ejector, 3-way syringe tip, suction filter ,spittoon
Hex
Socket
Hole
Connerer
Fuse
Water tank
Suction filter
Strong suction
Weak suction
Spittoon
Saliva ejectors
3-way syringe tip