-3-
III. SPECIFICATIONS
Refrigerant
Refrigerant
Cond. Unit Amp
Model No.
HP
Type
Charge
Draw (LRA)
Voltage
Amps
UHT27
1/6
R-134a
8 oz.
4.0
120/60/1
3.4
UHT48
1/4
R-134a
12oz.
6.1
120/60/1
6.7
UHT60
1/4
R-134a
12oz.
6.1
120/60/1
6.7
ULT27
1/3
R-134a
9 oz.
7.2
120/60/1
6.7
ULT48
1/3
R-404A
14 oz.
8.9
120/60/1
10.0
ULT60
1/2
R-404A
14 oz.
10.8
120/60/1
12.0
UPT276
1/5
R-134a
9 oz.
4.5
120/60/1
4.3
UPT279
1/5
R-134a
9 oz.
4.5
120/60/1
4.3
UPT488
1/4
R-134a
12 oz.
6.1
120/60/1
7.0
UPT4812
1/4
R-134a
12 oz.
6.1
120/60/1
7.0
UPT4818
1/4
R-134a
12 oz.
6.1
120/60/1
7.0
UPT6012
1/4
R-134a
12 oz.
6.1
120/60/1
7.0
UPT6018
1/4
R-134a
12 oz.
6.1
120/60/1
7.0
UPT6024
1/4
R-134a
12 oz.
6.1
120/60/1
7.0
IV. OPERATING DATA
Refrigerator Model: UHT27
Refrigerant
R-134a
R-134a
Ambient Temperature
70
°
F
100
°
F
Suction Presure
Start of Cycle
43 lb.
43 lb.
End of Cycle
15 lb.
15 lb.
Discharge Pressure
Start of Cycle
68 lb.
71 lb.
End of Cycle
87 lb.
145 lb.
Freezer Model: ULT27
Refrigerant
R-134a
R-134a
Ambient Temperature
70
°
F
100
°
F
Suction Presure
Start of Cycle
11 lb.
11 lb.
End of Cycle
0 lb.
1 lb.
Discharge Pressure
Start of Cycle
84 lb.
120 lb.
End of Cycle
116 lb.
164 lb.
VII. ELECTRICAL OPERATION (cont’d)
-20-
VII. b - SEQUENCE OF OPERATION:
REFRIGERATION
1 - Conditions:
a. Unit connected to correct voltage.
b. Condensing unit controlled by thermostat.
c. Thermostat set to desired temperature.
d. Evaporator fan powered.
e. Cabinet temperature below set point.
f. High limit thermostat closed (freezer).
g. Power to thermostat contacts through defrost
timer 1/4 (freezer).
h. Door perimeter heater energized (freezer).
i. Defrost timer motor energized (freezer).
2- Temperature Rises Above Set Temperature:
a. Thermostat contact close, powering compressor.
3- Set Temperature Is Met:
a. Unit returns to conditions in step 1.
VII. c - COMPONENT LOCATION:
VII. b - SEQUENCE OF OPERATION (cont’d):
DEFROST MODE (freezer)
1 - Unit Operating In Refrigeration Mode.
2 - Defrost Intiated By Defrost Timer.
a. Defrost timer contacts 1/4 open.
1) Power removed from evaporator fan.
2) Power removed from thermostat contacts.
b. Defrost timer contacts 1/2 close.
1) Power to defrost heater.
3 - Defrost Time Expires.
a. Defrost timer contacts 1/2 open.
1) Power removed from defrost heater.
b. Defrost timer contacts 1/4 close.
1) Power to evaporator fan.
2) Power to thermostat.
4 - Unit Returns To Normal Operation Until The Next
Defrost Cycle.
27” Models
48” & 60” Models