TRANSTEK TeleCuff Gen 2 Скачать руководство пользователя страница 9

16

5

15

4. After the measurement, it will repeat the data  
transmission steps as described on pape 14.

When the “TRIPLE MODE” buton is located on “ON”, the monitor is under the Triple
Measurement mode.

1.When the monitor is off , press the “START/STOP” button to turn on, and then it will 
complete the whole measurement automatically.

2.When the first measurement is done, it won’t show 
the readings. It will start the second measurement after
60 seconds.

3. Then it will repeat the steps to finish the third  
measurement. When three measurements are done, 
the LCD will display the average of 3 readings.

Tip:

If the measurement fails in the process of 3 munimes, it will report 

an error and repeat the current measurement. If the error 

occurs for 2 times in a row, the mode of 3 times measurement will be

automatically quit.

Tip:

You can press the “START/ STOP” button to stop the measurement  

any time.

MEASUREMENT

MEASUREMENT

B: Using Triple Measurement Mode

Display the zero point

Inflating and measuring

LCD display

Содержание TeleCuff Gen 2

Страница 1: ...0 Guangdong Transtek Medical Electronics Co Ltd Zone A No 105 Dongli Road Torch Development District 528437 Zhongshan Guangdong China User Manual Pro Cellular Blood Pressure Monitor TMB 2092 G TeleCu...

Страница 2: ...c pressure What is the standard blood pressure classi cation Why does my blood pressure uctuate throughout the day Is the result the same if measuring on the righe arm ABOUT BLOOD PRESSURE TROUBLESHOO...

Страница 3: ...vice Date of Manufacture Recyclable Indicates the need for the user to consult the instructions for use for important cautionary information such as warnings and precautions that cannot for a variety...

Страница 4: ...ency of washing skin condition and storage state The This device is intended for indoor use This device is portable but it is not intended for use during patient transport This device is not suitable...

Страница 5: ...nt icon loose when measuring will appear if the cuff is not wrapped or wrapped too Cuff wearing detection icon Indicates the current measurement mode is Triple Mode and which measurement progress it i...

Страница 6: ...e the AC adapter authorized by the manufacturer Please unplug the adapter to depart from the using utility power when you nish the measurement 6V 1A Do not use new and used batteries together Do not u...

Страница 7: ...w rst Note B turned off If unsuccessful the monitor will stop searching and enter the standby mode This device is equipped with function of Measurement Reminder Contact Customer Support to enable or d...

Страница 8: ...ill appear on the LCD Adjust the zero point In ating and measuring Display the measured result When the TRIPLE MODE button is located on OFF the monitor is under the Single Measurement mode When the m...

Страница 9: ...start the second measurement after 60 seconds 3 Then it will repeat the steps to nish the third measurement When three measurements are done the LCD will display the average of 3 readings Tip If the m...

Страница 10: ...dry place and avoid the sunshine Avoid intense shaking and collisions Using wet cloths to remove dirt Avoid washing the cuff Avoid touching water clean it with a dry cloth in case Avoid dusty and uns...

Страница 11: ...e range Please note that only a physician can tell whether your blood pressure value has reached a dangerous point Blood Pressure Category Normal Elevated High Blood Pressure Hypertension Stage 1 High...

Страница 12: ...K THIS REMEDY No power Low Battery Error message Display can not light up Batteries are exhausted Replace with new batteries Install the batteries or plug in the adapter properly Replace with new batt...

Страница 13: ...ressure 0mmHg 299mmHg SYS 60 mmHg 230 mmHg DIA 40 mmHg 130 mmHg Pulse value 40 199 beat minute Device Classi cation Expected Lifetime Battery Powered Mode Internally Powered ME Equipment AC Adaptor Po...

Страница 14: ...45 90 135 180 225 270 and 315 0 UT 1 cycle and 70 UT 25 30 cycles Single phase at 0 0 UT 250 300 cycle Conduced RF IEC61000 4 6 3 V 0 15 MHz 80 MHz 6 V in ISM and amateur radio bands between 0 15 MHz...

Страница 15: ...TE Band 13 17 Pulse modulation 217 Hz GSM 800 900 TETRA 800 iDEN 820 CDMA 850 LTE Band 5 Pulse modulation 18 Hz GSM 1800 CDMA 1900 GSM 1900 DECT LTE Band 1 3 4 25 UMTS Pulse modulation 217 Hz Bluetoot...

Отзывы: