This troubleshooting guide references devices manufactured during different time frames. It is important to 
know the manufacture date and serial number of your device during the troubleshooting process or when 
calling customer service. To determine the year of manufacture, refer to the first six (6) digits of your TMM 

Series Stretcher-Chair’s serial number – these represent the date the chair was manufactured. 
 

If you are utilizing an electronic copy of this document, words that are 

bold, blue and underlined

 

represent 

links to other pages in the document and/or links to web pages outside of this document. Simply click on the 
words to follow the links.  
 
Our Customer Service Department can be reached by any of the following methods:  
Email: 

[email protected]

 

Phone: 352-854-2929 or 800-237-3377 (Toll Free U.S. Only) 
Fax: 352-854-9544 

Hours of Operation: Monday – Friday, 8:30 am – 5:00 pm ET  
 

 

 

Содержание TMM Series

Страница 1: ...hair s serial number these represent the date the chair was manufactured If you are utilizing an electronic copy of this document words that are bold blue and underlined represent links to other pages...

Страница 2: ...g Page 5 a Follow this link if one or more portions of the device is working but some are not moving b If the Stretcher chair is an X Series chair the manufacture date is prior to February of 2012 and...

Страница 3: ...esting protocol can be found by following this link B Battery Testing Protocol For A AC Powered chairs ensure the device is plugged into a working 120V outlet AC powered chairs require working batteri...

Страница 4: ...eshoot the issue Both devices must have the same or similar control boxes 5 If switching pendants resolved the issue contact our Customer Service Department for a replacement pendant 6 If another pend...

Страница 5: ...ely failed and may need to be replaced If this is an X Series chair built prior to February of 2012 additional troubleshooting can be done by following this link X Series Chair built prior to February...

Страница 6: ...echnical service for part number and price 7 If the same section of the column is still not working and the chair has a KOM control box then the issue is likely in the column Call or email technical s...

Страница 7: ...mit switch appears to be working properly access the control box in the base of the chair and unplug the cord in port 4 WARNING This will disable the limit switch allowing the chair to be lowered whil...

Страница 8: ...d in the actuated position the tilt motor will not move in that direction 4 Using a Phillips head screwdriver remove the cover and attempt to manually actuate the switches 5 An audible click should be...

Страница 9: ...etermine the faulty part in the chair 3 Note which motor is currently functioning and which is not currently functioning 4 Unplug the motor that is not functioning and plug this motor into the port fo...

Страница 10: ...supply B Chairs Only Power supply A Chairs Only Port 10 Pendant pendant extension on X series chair Port 7 Limit Switch TMM3 Only Control Box Port Diagrams KOM control box diagram on next page VCU Co...

Страница 11: ...of 2012 Return to one of the following sections Tall T Chair is Not Going to Full Height One or More Functions Have Stopped Functioning or the General Issues page Pendant Port 1 Column Port 2 Back Mot...

Страница 12: ...Please refer to section 6 of Winco Mfg LLC s Terms Conditions for additional information concerning our return policy When returning parts please contact our Customer Service Department customerservi...