Document No.: OM TMM4 SERIES 

Revision: X 

Page 29 of 30 

 

PAD REMOVAL INSTRUCTIONS 

 

When removing cushions, pull laterally. Do not pull from top to bottom.

 

PREVENTIVE MAINTENANCE 

 
To ensure proper operation and extend the chair

’s life, it should be maintained on a regular basis as shown below. 

 

COMPONENT

 

FREQUENCY

 

PROCEDURE

 

TOOL

 

COMMENT

 

Cushion Pads 

After each use 

 

Inspect for 

tears. 

 

If torn, discard 
cushion. 

All Rails 

Every three months 

 

Inspect for 
chipped paint, 
burrs, rough 
edges. 

 

Remove burrs, 
rough edges. 

 

Touch up paint.  

 

Metal file 

 

Touch-up paint 

 

 

To prevent 
personnel cuts, 
burrs and rough 
edges should 
be removed 
once found. 

All Fastened Joints 

Every three months 

 

Inspect all 
fasteners to 
ensure proper fit 
and tightness. 

 

Retighten as 
needed. 

 

Wrenches 
(various sizes) 

 

Allen wrenches 
(various sizes) 

 

Screwdriver 

 

All Labels 

Every three months 

 

Inspect for 
tattered, torn, 
missing, and 
illegible labels. 

 

Call TMM 
Service for new 
labels. 

Pendants 

Every three months 

 

Test function of 
each button  

 

Inspect pendant 
label to ensure it 
is readable.  

 

Call TMM 
Service if 
pendant is not 
functioning or 
label is 
unreadable  

Содержание TMM4 Series

Страница 1: ...del TMM4 Series Multi Purpose Stretcher Chair Options W Wide T Tall R Rail on Back Section Q Scale L Low G ENG Headpiece F Folding Footrest E Export B Battery A AC Power 5516 S W 1st Lane Ocala FL 344...

Страница 2: ...12 PENDANT CONTROLLER 15 PATIENT INGRESS EGRESS 16 PATIENT TRANSFER 17 PUSH BAR OPERATION 17 OPTIONS 18 W OPTION WIDE WIDTH 18 T OPTION TALL HEIGHT 18 R OPTION RAILS ON BACK SECTION near patient s hea...

Страница 3: ...sted are nominal and operation may vary slightly from unit to unit due to tolerance variations or power supply variations Photographs and illustrations contained within this document may not depict ex...

Страница 4: ...ng battery to open flames cigarettes sparks and incandescent materials Never charge battery in enclosed unventilated spaces Do NOT store battery in sealed container Store in fresh well ventilated area...

Страница 5: ...se unpredictable operation resulting in injury to patient medical personnel or bystander Modifying chair will void warranty USE SAFE OPERATING PROCEDURES Prior to operating chair ensure patient cleara...

Страница 6: ...ist EMI up to a certain intensity called its immunity level When the immunity level is higher the less likely EMI will interfere with chair operation The chair has been tested to 20 V m immunity level...

Страница 7: ...Installation and Usage Instructions IM TMA57 15 under the Cautionary Usage Instructions section for more information For A Option Only If chair is to be stored or not in use for more than 1 5 weeks l...

Страница 8: ...Document No OM TMM4 SERIES Revision X Page 8 of 30...

Страница 9: ...Document No OM TMM4 SERIES Revision X Page 9 of 30...

Страница 10: ...SHOULD BE MOVED TO A MORE STABLE NON PADDED PLATFORM SECONDARILY IF ATTEMPTING TO PERFORM CHEST COMPRESSIONS ON THE TMM DEVICE THE PATIENT SHOULD BE PLACED ON A BACK BOARD OR POSITIONED ON THE CHAIR...

Страница 11: ...s from spinning NEUTRAL MODE Activate neutral mode by placing either brake pedal into a horizontal orientation This mode allows all four casters to swivel and all wheels to spin freely STEER MODE Acti...

Страница 12: ...depending on rail location Push side rail until it fully engages retainer clip RAISING SIDE RAIL Lift side rail until it is in up position Red rail release tab will engage lock automatically Ensure s...

Страница 13: ...Document No OM TMM4 SERIES Revision X Page 13 of 30 To move inward rotate bottom jam nut as shown To move outward rotate bottom jam nut as shown...

Страница 14: ...aligned top nut is tightening nut 2 When adjusting ensure bottom jam nut is contacting seat frame 3 Ball joint head angle does not matter 4 Prior to adjustment pull outward on side rail in order for s...

Страница 15: ...o lower chair back LEG SECTION ADJUSTMENT Second row of pendant buttons controls angle of leg section Press left button to raise chair s leg section Press right button to lower chair s leg section SEA...

Страница 16: ...ter wheels 2 Adjust chair to lowest height and into upright chair position 3 Lower one side rail of seat section 4 If back section rails are present remove lower back section rail of same side 5 Posit...

Страница 17: ...into horizontal orientation Press Auto Contour button 2 Lower remove all side rails from transfer side of chair 3 Position chair as close as possible to other surface 4 Match chair stretcher height to...

Страница 18: ...OM TMM4 SERIES Revision X Page 18 of 30 OPTIONS W OPTION WIDE WIDTH For W option patient surface width is 28 Standard width is 24 T OPTION TALL HEIGHT For T option chair height travel is 16 Standard...

Страница 19: ...Document No OM TMM4 SERIES Revision X Page 19 of 30 R OPTION RAILS ON BACK SECTION near patient s head Rails located on back section operate and adjust same as side rails...

Страница 20: ...measurement tare scale before each patient To prevent inaccuracy prevent extra items accessories monitors IV bags blankets pillows charts etc from being included excluded during weight checks To ensu...

Страница 21: ...utton is pressed 1 Remove battery cover on back of pendant by sliding it downwards 2 Take out battery and replace it with a new 9V DC Industrial battery 3 Replace battery cover To ensure proper scale...

Страница 22: ...or over a bump G OPTION ENG HEADPIECE To remove ENG headpiece 1 Loosen two black knobs on either side of chair 2 Lift headpiece from back section of chair To prevent component loss ensure that knobs r...

Страница 23: ...mfort stow footrest prior to articulating chair into supine stretcher position EXTENDED FOOTREST STOWED FOOTREST E OPTION EXPORT E option designates chair was shipped with power cord specific to recei...

Страница 24: ...CHAIR Place two round cam locks on back of battery into square openings of mounting bracket Once in place slide battery down until cam locks engage ADDITIONAL BATTERY INFORMATION All chairs come with...

Страница 25: ...batteries overnight i e during off times or 24 hours prior to service When the chair is plugged in and charging the pendant will not illuminate even when buttons are pushed If chair is to be stored or...

Страница 26: ...OUND WIRE ATTACHES TO GROUND POST A OPTION ONLY BATTERY CORD PLUGS INTO PORT 13 COLUMN PLUG TMM4 OR PLUGS TMM4 T PLUG INTO PHONO TO DIN ADAPTER PLUGS THEN PLUG INTO PORTS 1 AND PORT 2 FOR TMM4 T BACK...

Страница 27: ...els Actuators and Battery Pack As required in accordance with facility policies Wipe with damp cloth and allow to air dry Facility approved detergents disinfectants water Clean exterior surfaces only...

Страница 28: ...ry should be cleaned on a regular basis with a damp cloth soaked in a mild soap and water solution Avoid harsh detergents or chemicals that could damage the finish of your product If disinfecting with...

Страница 29: ...ed paint burrs rough edges Remove burrs rough edges Touch up paint Metal file Touch up paint To prevent personnel cuts burrs and rough edges should be removed once found All Fastened Joints Every thre...

Страница 30: ...ich can cause battery explosion or fire To request Material Safety Data Sheet MSDS for battery contact TransMotion Medical Customer Service BATTERY LIFE Battery capacity diminishes over time Charge ba...

Отзывы: