Transmille EA008 Скачать руководство пользователя страница 6

EA003 OPTICAL TACHOMETER CALIBRATION ADAPTER 

OPERATION MANUAL 

6: Connect current and turn on. 
 
Note 1:  EMC interference and pick up can make low current readings unstable. Great care must be  
Taken to screen all leads and connections for stable readings. 
 
Note 2:  The BMC outer shell will be earthed if the calibrator is set  ‘negative to ground’. 
This may cause earth loop errors is the current source is also earthed. 
 

Measure High Value resistance. 

1: Connect to calibrator as above. 
2: Start the VFP program, select com port as required. 
3: Click ‘resistance’ button 
4: Click on resistance range required or the ‘Auto range’ button. 
5: With inputs open circuit click the ‘Set infinity’ button. 
7:  Using the drop down box select the voltage to measure the resistance at. 
6: Connect the resistance to be measured between the ‘voltage supply’ BMC socket and the ‘Input 
current’ BMC socket. Note the outer shield of the BMC is ground to connect to the screen/guard. 
7: Click ‘Volts on’ button 
8: Read value on display. 
 
Note 1:  EMC interference and pick up can make High Resistance readings unstable. Great care 
must be taken to screen all leads and connections for stable readings. 
 
WARNING: 
The pAmp input of the calibrator is very sensitive and can be damaged by the application of 
voltages above 50V. 
 
 

 
 
 

Detailed  VFP Operation 

 

Selecting a COM Port 

 

 

 
On starting the virtual front panel for the first time, a COM port selection screen will be 
displayed. Click on the COM port which is connected to your 2000 Series calibrator to 
continue. 

Transmille Ltd.  

Page 6 

Содержание EA008

Страница 1: ...EA008 High Resistance pA Measurement Adapter Operation Manual Version 1 00 April 2007 All product names are trademarks of their respective companies...

Страница 2: ...nsmille will contact the purchaser before proceeding with any repair To obtain repair under this guarantee the purchaser must send the instrument in its original packaging carriage prepaid and a descr...

Страница 3: ...DAPTER 4 CONNECTIONS TO THE 3000 SERIES CALIBRATOR 5 LOADING THE VFP SOFTWARE ONTO PC 5 BASIC OPERATION STEPS AND CONNECTION TO UUT 5 Measure pA 5 Measure High Value resistance 6 DETAILED VFP OPERATIO...

Страница 4: ...ent measurement ranges down to 10nA with a resolution of 1pA The EA008 provides a virtual ground input impedance avoiding errors due to high input impedance often found on DMM s with low current range...

Страница 5: ...t panel of the 3000 series calibrator The calibrators RS232 serial interface must also be connected to a PC running the VFP program Loading the VFP software onto PC Insert the CD supplied into the PC...

Страница 6: ...nfinity button 7 Using the drop down box select the voltage to measure the resistance at 6 Connect the resistance to be measured between the voltage supply BMC socket and the Input current BMC socket...

Страница 7: ...selection buttons Sampling is indicated on the main display by Functions Current Measurement The current measurement function provides five ranges to measure current 100uA 10uA 1uA 100nA 10nA Functio...

Страница 8: ...resistance in green with a smaller display of current in blue Auto Range Function Use the AUTO button to turn on the autoranging function Click AUTO again or select a range button to turn off autorang...

Страница 9: ...nals should be OPEN CIRCUIT If auto range is selected the virtual front panel will step through each range and set infinity Care Maintenance The only maintenance instructions for the adapter is period...

Страница 10: ...shock or be dropped Transportation is recommended using the original packaging with avoidance of extreme changes of temperature Servicing Information The adapter is provided certified from the factory...

Страница 11: ...OMETER CALIBRATION ADAPTER OPERATION MANUAL EA008 High Resistance pA Measurement Adapter Appendix A Specifications Ranges 10nA 100nA 1uA 10uA 100uA Resolution 1pA 10pA 100pA 1nA 10nA Accuracy 0 5 Tran...

Отзывы: