The EIA/TIA 568 standard for LAN cables only uses 4 

of the wires in the eight-wire cable. The other 4 wires 

in the cable may have faults, but these will not affect 

the operation of the LAN. The CNT 10 L tests all of the 

wires in the LAN cable, and identifies the faults. Even 

though these wires may not being used in the LAN 

system.
Not all LAN cables are wired straight through Crossover 

cables used on LANs are purposely miswired, with their 

Receive and Transmit wires cross over. The CNT 10 L 

will test this as a bad cable, but it may work just fine as 

Crossover cable.

7.6.2 The CNT 10 L says my cable is good, but it 

does not work on my LAN

: Many cable testers like 

the CNT 10  L only perform continuity style tests (open, 

short. crossed, etc). Ethernet LAN cables are const-

ructed in a special way. The 8 wires inside are grouped 

into 4 pairs of 2 wires each. Not only must the 8 wires 

connect form end to end of the cable, the pairs in the 

cable must connect to specific pins of RJ-45 plugs on 

to the ends of cable, ignoring the pairing(as described 

in ElA/TIA 568), and the cable will test ok. But when 

the cable is tried on the LAN, it does not work. This is 

because the Iack of the proper pairing causes exces-

sive cross talk in the cable. Preventing the LAN from 

working. This type of cable fault is sometimes call a split 

pair or double split pair, to detect this type of fault, a 

more sophisticated tester capable of performing a NEXT 

test (Near End Cross Talk) must be used. 

Notes:

 

Cross talk increases with the length of the cable. A LAN 

system will tolerate a certain amount of cross talk. A 

short cable (10ft or less) that is improperly paired may 

work just fine. However, Ionger cables, paired in exactly 

the same way, may not work. This explains why an 

installer can make short Jumper cables that work (al-

though they are improperly paired), but when he installs 

RJ-45‘s in exactly the same way on a Ionger cable, the 

cable does not work.

7.6.3 The CNT 10 L says my phone cable is bad, but 

my 

phone works oK.

 

Most single line telephones only use 2 wires in a 

modular cable. The modular cable, which terminates in 

RJ-11connectors, could have as may as 6 wires in it. 

The unused wires may have faults, which the  

CNT 10 L identifies, but these faults may have no effect 

an the working 2-wire telephone circuit.

CNT10L_Manual_en.indd   11

25.06.2019   14:19:13

Содержание CNT 10 L

Страница 1: ...Network Cable Tester Manual CNT 10 L CNT10L_Manual_en indd 1 25 06 2019 14 19 12...

Страница 2: ...es The 4 in 1 Cable Tester tests 4 common LAN and Computer cables It tests installed cables or patch cords with RJ 45 RJ 11 USB and BNC connectors It is intended to test cables with straight through c...

Страница 3: ...first disconnecting it from all external circuitry Failure to observe this precaution may result in damage to the tester or injury to the user 2 6 Repairs and maintenance must only be carried out by q...

Страница 4: ...sing 2 14 Do not modify this tester Changing the design may make the tester unsafe and may result in injury to the user 2 15 Do not use this tester if it has undergone long term storage under unfavora...

Страница 5: ...ded STP or unshielded UTP LAN cables 3 12 Test shields in USB cables 4 1 Cables Tested UTP and STP LAN cables Terminated in RJ 45 male connectors ElA TIA 568 RJ 11 cables with male connectors 2 to 6 c...

Страница 6: ...Contents 6 5 1 Test Button 5 2 Main Display 5 3 RJ 45 Jack 5 4 RJ 11 Jack 5 6 USB Jack 5 7 BNC Jack CNT10L_Manual_en indd 6 25 06 2019 14 19 13...

Страница 7: ...The Remote is often attached to the Main unit when storing when transporting or when patch cables are being tested The Remote is removed from the Main unit when an installed in wall ceiling etc cable...

Страница 8: ...with light and the Beeper will sound once 6 6 4 Connected LED Lo Hi Beep Numbered LEDs If the Connected LED lights and the Beeper emits a LoHi beep examine the numbered LEDs i e LEDs numbered 1 to 8 a...

Страница 9: ...n crossed This seldom affects the performance of Standard analog telephone lines POTS Digital telephone lines and old touch tone phones may be polarity sensitive so a cros sed cable may prevent them f...

Страница 10: ...p or put in ser vice and it tests as Open or Crossed the RJ connec tors have probably been installed incorrectly Shorts very seldom occur as the result of a badly crimped RJ connector so the user shou...

Страница 11: ...e cable will test ok But when the cable is tried on the LAN it does not work This is because the Iack of the proper pairing causes exces sive cross talk in the cable Preventing the LAN from working Th...

Страница 12: ...e line rings 7 6 4 The CNT 10 L says my RJ 11 phone cable is good but my phone doesn t work First make sure the telephone line is active The good cable may result in a reversed line The line is not ac...

Страница 13: ...justments Calibration is not required To clean the outside of the tester use a cloth dampened with a mild detergent solution Do not use any abrasive cleansers or chemical solvents that may damage the...

Отзывы: