Transforming Technologies Ohm Metrics EFM115 Скачать руководство пользователя страница 9

 

 

Operation:  Field Meter 

1.  The field meter must be grounded properly 

with the ground lead provided. 

2.  Press the 

Power

 Button once to activate 

the meter. A single beep will indicate 
proper operation.   

3.  The meter reading should be zero (0).  

Use the 

Reset

 button if it is not. 

4.  Hold the field meter 1" away from a surface, using the 

"Focus Rings" for proper distance. 

5.  A bar graph and a numerical LED readout displays the 

voltage of the charged surface. Press 

Hold

 to freeze 

the screen to better read the result. 

6.  Press the red 

Power Button 

to turn off meter.  

 

Understanding the Results 

Easy-to-read LCD Display 

The LCD Display on the EFM115 is divided into three sec-
tions. The upper left section displays a bar graph in blue 
to indicate a negative polarity.  The upper right section 
displays a bar graph in red to indicate a positive polarity.  
The lower section display numerical values, voltage 
measurement and function sections. 

For the negative polarity voltage 

Functions in use 

Battery Condition 

Measurement Unit 

Voltage 

For the positive polarity voltage 

Содержание Ohm Metrics EFM115

Страница 1: ...on Manual Static Field Meter Ionizer Verification Unit Model EFM115 T R A N S F O R M I N G T E C H N O L O G I E S L L C O U T S T A N D I N G A L T E R N A T I V E S I N S T A T I C C O N T R O L Oh...

Страница 2: ...M115 1 Ohm Metrics 1 Specifications 2 Operation Battery Installation 2 Static Field Meter 2 Ion Balance 2 Measuring Two Point Resistance 3 LED Indication Guide 4 Calibration 4 3 Parts Included List 4...

Страница 3: ...meter and ion balance tester makes the EFM115 the ideal meter for static audits and periodic verification of ionizers and meets periodic test requirements per Compliance Verification ESD TR53 and con...

Страница 4: ...may be disabled if preferred Product Specifications Product Number EFM155 Electrostatic Field Meter Measuring range Static voltage 0 1 49 kV Low range 1 0 20 0 kV High range Ion balance voltage 0 200...

Страница 5: ...removable cover marked OPEN NOTE Be careful while connecting or disconnecting the battery To avoid breaking the contacts do not pull the connection plug hard 1 Press down on OPEN and slide the cover...

Страница 6: ...D lights The below message appears during sensor failure Four colorcoded buttons to switch Measurement modes Unitcomes with handy carrying case Openthe lid onthe back ofmeter to change battery Ground...

Страница 7: ...of inactivity the unit will beep 5 times and turn itself off To resume operations press the Power button once The auto off feature can be disabled by pressing and holding the Power button for at leas...

Страница 8: ...uipped with two Focus Rings that are beamed from the front of the meter in ensure proper measuring distance Two beams of light will form con centric circles when the meter is the correct 1 away from t...

Страница 9: ...he charged surface Press Hold to freeze the screen to better read the result 6 Press the red Power Button to turn off meter Understanding the Results Easy to read LCD Display The LCD Display on the EF...

Страница 10: ...atic charge may damage the sensor Operation Ion Balance The compact and lightweight design of the EFM115 makes it ideal for periodic verification of ionizers The EMF115 can be used to measure low volt...

Страница 11: ...e such as one foot 7 A bar graph and a digital display on the EFM115 display the ion balance information Press Hold to freeze the display 8 When finished with Ion Balance Mode depress the charge plate...

Страница 12: ...gies LLC provides a limited war ranty for the Model EFM115 All new products are guaranteed to be free from defects in material and workmanship for a pe riod of one 1 year from the date of shipment Lia...

Отзывы: