Breathing circuit not
assembled correctly
Replace any external hardware
exceeding recommended service
life.
If the problem persists, contact
your homecare provide.
All LEDs flash four
times then pause,
then flash again
Device detects an
operating error.
Hold down the power button until
all LEDs stops flashing. Release
the power button, the device will
reset and power up in Standby
Mode. If the LEDs begin to flash
again, contact your homecare
provider.
Droplets of water in
my mask after therapy
Excess condensation from
breathing
Run drying mode after therapy.
Utilize in a less humid
environment (dehumidifier).
Bluetooth will not
connect
Bluetooth disable (White
LED light)
Push Bluetooth button to enable
(Blue LED light).
App not downloaded or
issue with the app/phone
function
Verify app is downloaded and
installed, the phone supports
Bluetooth Low Energy and has
Bluetooth enabled.
SleepStart is not
functioning
Breathing circuit has
excessive leaks
Check hose, mask and any
accessories in the breathing
circuit for air leaks and proper fit.
User not breathing deep
enough to trigger
SleepStart
Ensure deep breath is taken with
mask securely fitted or use press
Power button to start therapy.
If the problem persists, call your
homecare provider’s technical
service department.
Precautions for Use
This section describes the warnings and cautions associated with use of the Transcend
Micro. The following guidelines apply to this document:
Warning
Indicates the possibility of serious injury or death to yourself or others.
Caution!
Indicates the possibility of minor injury or damage to the equipment.
NOTE:
Indicates a tip, explanation, or feature to aid in understanding, or efficient
operation of the device.