background image

12 

 

17.  In  order  to  save  all  parameter  settings,  it  is  imperative  to  exit  all  menus  in  the  following 

manner: At the parameter level, press the UNITS (up) key twice (two times) to exit the User 
Menu. The instrument displays ‘SAvE’ and then automatically powers off.   

 

Diagnostics (A24) 

 

Here is a brief description of each test mode: 
 

A24-U1  Display  Test

  –  Lights  up  all  display  segments,  counting  down  from  666666  to 

11111. Test ends automatically when complete. 

A24-U2  ADC  Test

  –  Shows  internal  A/D  converter  counts  –  useful  for  troubleshooting 

weighing issues. End test manually by pressing the NET/GROSS (Set) key.  

A24-U3  Input  Test

  –  Displays  input  logic  (0  or  1)  of  input  terminal.  “0”  mean  input  pin  is 

grounded;  “1”  means  input  pin  is  open.  End  test  manually  by  pressing  the  NET/GROSS 
(Set) key. 

A24-U4  Output  Test

  –  Sets  all  output  pins  ‘ON’.  End  test  manually  by  pressing  the 

NET/GROSS (Set) key. 

A24-U5 Serial Test

 – Transmits a data string continuously out both serial ports (“TEST1” on 

COM1 and “TEST2” on COM2). End test manually by pressing the NET/GROSS (Set) key. 

 

A24-U6  Keyboard  Test

  –  Displays  a  keycode  for  each  key  pressed  on  the  keypad.  See 

Table below. End test manually by pressing the NET/GROSS (Set) key. 

 

Key 

Keycode 

UNITS 

ZERO/OFF 

NET/GROSS 

TARE 

ON/PRINT 

 

Entering the Bluetooth (“B”) Menu

 

 

NOTE

: This menu applies to remote wireless A/D systems only. 

 

1.  Switch  off  the  display  unit  by  pressing  and  holding  down  the  ZERO/OFF  key  for  about  5 

seconds. 

2.  Press  and  hold  down  the  ON/PRINT  key  (about  20  seconds)  until  the  screen  shows  

“-F-”. 

3.  Press the ON/PRINT key until the screen displays “-B-”. 

4.  Scroll down using the ZERO/OFF (down) key to reach the parameter level. The instrument 

shows “B   1”. 

5.  Move from one “B” parameter to the next by using the TARE (left) or ON/PRINT (right) keys. 

For  example,  to  go  from  B1  to  B2,  press  the  ON/PRINT  key.  To  go  from  B2  back  to  B1, 
press the TARE key. 

6.  Once  you  have  arrived  at  the  proper  “B”  menu  parameter,  e.g.  ”B    1”,  press  the  

ZERO/OFF (down) key once to arrive at the selection level. The instrument displays the cur-
rent parameter setting. 

Содержание TI-500 FDI Series

Страница 1: ...MODEL TI 500 FDI Force Data Instrument Installer s Manual Revision 1 1 October 14 2016...

Страница 2: ...OF TI 500 FORCE DATA INSTRUMENT 2 GETTING STARTED 3 SYSTEM CONFIGURATION 4 CONFIGURATION MENUS 4 ENTERING THE SETUP F CONFIGURATION MENU 5 SETUP F MENU DESCRIPTIONS 5 ENTERING THE USER A OR COM C MENU...

Страница 3: ...libration of the force data instrument For operation and troubleshooting please refer to the separate user s guide Installation of TI 500 force data instrument Find a suitable location for the instrum...

Страница 4: ...batteries to operate not included To install the batteries remove the plastic battery cover from the rear panel Observe proper direction polari ty of the batteries Replace the battery cover The TI 500...

Страница 5: ...NOTE This menu will not appear for cabled systems see F29 The configuration menus are laid out in the following vertical arrangement Top Menu selection level Parameter level Selection level or functi...

Страница 6: ...meter settings it is imperative to exit all menus in the following manner At the parameter level press the UNITS up key twice two times to exit the Setup Menu The instrument displays SAvE and then aut...

Страница 7: ...re Press ZERO OFF key to begin se quence F16 Zero Calibra tion Places instrument into the zero calibration routine Scrolling down with the ZERO OFF key one level begins the procedure Press ZERO OFF ke...

Страница 8: ...e AdC Internal A D cabled 1rAdIo One external wireless A D module 2rAdIo Two external wireless A D modules AdC 1rAdIo 2rAdIo F30 Special Appli cation Used to select one special application feature sub...

Страница 9: ...o go from A1 to A2 press the ON PRINT key To go from A2 back to A1 press the TARE key 6 Once you have arrived at the proper A menu parameter e g A 1 press the ZERO OFF down key once to arrive at the s...

Страница 10: ...bit 7_o_2 7 data bits with odd parity bit and two stop bits 8_n_1 8_n_2 7_n_1 7_n_2 7_E_1 7_E_2 7_o_1 7_o_2 A3 C3 Serial Port Mode Selects the mode of the serial port Refer to Appendix B for more info...

Страница 11: ...lects the printed format for date USA mm dd yy IntL dd mm yy USA IntL A20 Set System Time Date Allows you to set the system time and date Pressing the ZERO OFF key to scroll down one level begins the...

Страница 12: ...ashing digit 10 After entering the exact value press the NET GROSS key to save the value The screen displays dA_xx where xx is the current day of the month e g 14 One digit will be flash ing 11 Use th...

Страница 13: ...uously out both serial ports TEST1 on COM1 and TEST2 on COM2 End test manually by pressing the NET GROSS Set key A24 U6 Keyboard Test Displays a keycode for each key pressed on the keypad See Table be...

Страница 14: ...one level begins the programming sequence Key in 0 to 6 5535 1 0000 B3 Adjust Load Cell Corner 3 Factor This parameter allows you to adjust tweak the digital corner com pensation factory for load cell...

Страница 15: ...s only 1 Switch off the force data instrument by pressing and holding the ON PRINT key for about 5 seconds 2 Press and hold down the ON PRINT key about 20 seconds until the screen shows F 3 Press the...

Страница 16: ...the calibration was successful the display will show EndC1 momentarily followed by C 2 for the second calibration point 9 Repeat steps 7 8 for C2 thru C7 At the conclusion of C7 the instrument displa...

Страница 17: ...r in the actual positive rated capacity value 3 After setting the exact value press the NET GROSS Set key to save the value The indicator will prompt you to enter the positive span in mV V P S 4 Use t...

Страница 18: ...ollowing in sequence and then returns to the F CAL level Span Point 1 Weight Value Span Point 1 A D Count Span Point 2 Weight Value Span Point 2 A D Count Span Point 3 Weight Value Span Point 3 A D Co...

Страница 19: ...on is transmitted continuously making it a popular choice for remote displays and other remote devices requiring a constant data stream The trans mission automatically occurs at the end of each displa...

Страница 20: ...ed if it is disabled in A18 of the User Menu STRING FORMAT 1 Condec Demand String String Format 1 is designed for two way communication STX POL xxxxx xx LB KG GR NT CR LF Polarity SP Positive Negative...

Страница 21: ...ogram Gross Net G Gross N Net Start Transmission Carriage Return Line Feed Status SP Valid M Motion O Over under range P pieces C custom units PCS pieces NOTE Additional unit is available N Newtons O...

Отзывы: