8. BASIC OPERATIONS
5
6
Prepare suitable mass standard.
Load the weighing pan with the mass
standard and press key.
Prior to adjustment prepare the right mass
standard. You are recommended to use a mass
standard of class F1 or F2. View the table and
select the mass standard you need.
8.1. Adjustment
External adjustment is carried out using an external mass
standard of the right accuracy and weight value, which
value depends on balance type and capacity.
Balance
model
TCS-203
200
TCS-602
500
TCS-2002
TCS-3101
2 000
2 000
Mass
standard [g]
1
2
Simultaneously press +
keys to enter
home screen.
While “P1.CAL” parameter is on, press key.
3
4
Unload the weighing pan and press key.
While "1.1.CA-E" parameter is on, press
key.
Содержание PROCISION TCS Series
Страница 1: ...Start up Guide Series Balances TCS Transcat com 800 828 1470 ...
Страница 10: ...9 PERIPHERAL DEVICES CONNECTORS ...
Страница 12: ...Transcat com 800 828 1470 ...