13
8.12
While temperature digit is blinking. Press the
SELECT
button
up
or
down
to adjust the value to that of the solution.
8.13
Press
ENTER
to confirm the set value or
EXIT
to abort. Display will
resume measuring mode.
9
MAKING REDOX(ORP) MEASUREMENT:
9.1
Disconnect and remove the pH electrode and temperature probe.
Connect with a Redox/ORP electrode. Temperature has no effect
on ORP readings and therefore is not require during measurement.
9.2
Rinse the sensor in distilled water then in the test liquid. Stir to
remove bubbles and wait for a stabilized reading.
9.3
Press the
MODE
button once to switch display to
mV
reading.
9.4
Press the
READ
button once and display will show
AutoLock
symbol
blinking, indicating meter is sensing an endpoint reading.
9.5
When the
AutoLock
sign stops with short beep, an endpoint reading
is established. You can now record the reading.
9.6
Depend on the liquid that is measured, some volatile liquid may
take up to 20 minutes or even longer time to stabilize. Disable the
Auto-lock (Page4) and Auto-Off (Page6) features. Then endpoint
reading should determine by individual experience in each case.
DATA MEMORY:
9.7
This meter can save up to 99 data with time and
clock with either pH or ORP measurements.
STORE MEMORY:
9.8
After a reading has established, press the
M+
.
button
to save data into memory. Display will
show
Stor
Stor
Stor
Stor
and memory counter blink briefly.
9.9
Data is stored and measuring mode resumes.
9.10
When memory is above 99, display will show
“FULL”
and data will not be saved. Please
download data and clear memory to resume.
9.11
It is a good practice to download data daily.
RECALL MEMORY:
9.12
To review the saved data, press the
SELECT
button then press
RECALL
button.
9.13
Press the
SELECT
button
up
or
down
to review
each saved data. Time and date will alternate.
9.14
When finish reviewing, press
EXIT
to return to normal measuring
mode.
pH
7.08
25
.
0
Memory
2 8:
:
47
AM
RECALL
SELECT
EXIT
stor
Memory
2