8.12

 

While temperature digit is blinking.  Press the 

  SELECT 

 button 

up

 or 

down

 to adjust the value to that of the solution. 

8.13

 

Press 

 ENTER 

 to confirm the set value or 

 EXIT 

 to abort.  Display will 

resume measuring mode. 

9

 

MAKING REDOX(ORP) MEASUREMENT: 

9.1

 

Disconnect and remove the pH electrode and temperature probe.  

Connect with a Redox/ORP electrode.  Temperature has no effect 
on ORP readings and therefore is not require during measurement.  

9.2

 

Rinse  the  sensor  in  distilled  water  then  in  the  test  liquid.    Stir  to 

remove bubbles and wait for a stabilized reading. 

9.3

 

Press the 

 MODE 

 button once to switch display to 

mV

 reading. 

9.4

 

Press the 

 READ 

 button once and display will show 

AutoLock 

 

symbol 

blinking, indicating meter is sensing an endpoint reading. 

9.5

 

When the 

 AutoLock 

 sign stops with short beep, an endpoint reading 

is established.  You can now record the reading. 

9.6

 

Depend  on  the  liquid  that  is  measured,  some  volatile  liquid  may 

take up to 20 minutes or even longer time to stabilize.  Disable the 
Auto-lock  (Page4)  and  Auto-Off  (Page6)  features.    Then  endpoint 
reading should determine by individual experience in each case. 

 

DATA MEMORY: 

9.7

 

This meter can save up to 99 data with time and 

clock with either pH or ORP measurements. 

 

STORE MEMORY:

 

9.8

 

After  a  reading  has  established,  press  the 

    M+ 

.

  

button

 

to  save  data  into  memory.    Display  will 

show 

Stor

Stor

Stor

Stor

 and memory counter blink briefly. 

9.9

 

Data is stored and measuring mode resumes. 

9.10

 

When  memory  is  above  99,  display  will  show 

“FULL”

  and  data  will  not  be  saved.    Please 

download data and clear memory to resume. 

9.11

 

It is a good practice to download data daily. 

 

RECALL MEMORY:

 

9.12

 

To  review  the  saved  data,  press  the 

  SELECT 

 

button then press 

 

RECALL

 

 button. 

9.13

 

Press  the 

  SELECT 

  button 

up

  or 

down

  to  review 

each saved data.  Time and date will alternate. 

9.14

 

When  finish  reviewing,  press 

  EXIT 

  to  return  to  normal  measuring 

mode. 

 
pH

 

     7.08

 

 

    

25

.

0

 

 Memory 

   2   8:

 

:

47

 AM

 

 

       

RECALL 

 

SELECT 

EXIT 

 

 stor

    

    

  

 Memory 

  2   

 

 

Содержание WalkLAB HP 9010

Страница 1: ...0 15 0 11 4 1 14 5 10 13 2 9 6 3 12 7 8 15 0 15 0 1 14 1 14 13 2 13 2 3 12 3 12 11 4 11 4 5 10 5 10 9 6 9 6 7 8 7 8 WalkLAB Professional pH Meter HP 9010 Operations Manual...

Страница 2: ...2 online data logging to PC sold separately 2 CONTENT Carefully unpack the box It should contain the following items a Main unit b Combination pH electrode c Temperature probe d Calibration buffer sol...

Страница 3: ...to snap down cover 4 PRODUCT LAYOUT pH Electrode Socket Temperature Probe Socket Battery Cover Lift up latch here to open Battery Cover On Off Button Press once to switch ON To switch OFF press and h...

Страница 4: ...and Meter mS ppt ppm kPa ION DO Cond Salinity TDS S ATC Auto Lock AM PM HOLD pH Button keys legend correlate to button key Indicator will blink while the function mode is active Date or Time display L...

Страница 5: ...tomatically sense a stabilized reading and locks the endpoint reading Factory preset is Yes Yes Yes Yes 1 Press the ENTER key to set display will show Yes Yes Yes Yes blinking 2 Press the SELECT key u...

Страница 6: ...last calibration information about the electrode conditions A good pH electrode should have an Offset value no greater than 60mV and a Slope value of above 75 and below 115 Otherwise the electrode sh...

Страница 7: ...ELECT key up or down to choose between C C C C and F F F F 3 Press the ENTER key to confirm 4 Press the SELECT key up for next setting Avto off AUTO OFF TIME SETTING This setting allow you set the tim...

Страница 8: ...ENTER clock is up to date 5 Press the SELECT key up for next setting RST MASTER RESET This setting allows you to reset the unit to the original factory s default Original factory default Buffer Stand...

Страница 9: ...current temperature Calibration with ISO Standard 7 5 Make sure the unit is properly installed and electrode connected 7 6 This meter can automatically recognize the ISO Standard buffer solution You w...

Страница 10: ...bration cannot proceed until they are rectified 7 11 When the meter beeps Offset point is established 7 12 Now 1 68 4 01 10 01 and 12 45 will alternate on main display 7 13 Rinse the pH electrodes and...

Страница 11: ...uffer solution Temperature C Slope 1 68 Slope 4 01 Offset 6 86 Slope 9 18 Slope 12 45 0 1 67 4 01 6 98 9 47 13 43 5 1 67 4 00 6 95 9 38 13 21 10 1 67 4 00 6 92 9 32 13 00 15 1 67 4 00 6 90 9 27 12 81...

Страница 12: ...press EXIT for 2 seconds will abort all Calibration with Custom Standard 7 38 Make sure the unit is properly installed and electrode connected 7 39 Rinse the pH electrodes and temperature probe in dis...

Страница 13: ...t reading is established You can now record the reading 8 5 Press the READ button for each new measurement Measurement with Auto Lock feature disabled see Page 4 8 6 When feature is disabled meter wil...

Страница 14: ...Depend on the liquid that is measured some volatile liquid may take up to 20 minutes or even longer time to stabilize Disable the Auto lock Page4 and Auto Off Page6 features Then endpoint reading shou...

Страница 15: ...asure liquid within specified range ErLc Prob Readings drift or unstable cannot auto lock during Read or Cal 1 Ground loop in solution 2 Electrode is clog or expired 3 Electromagnetic interference 4 O...

Страница 16: ...nsinstruments com http www transinstruments com Singapore Pte Ltd email sales transinstruments com Website www transinstrumnets com ISO9001 Certified Firm Rev 4 Sep2015 PRINTED IN SINGAPORE Quality ch...

Отзывы: