USER GUIDE
21
RH Hood Method
For the best results ensure
surface is cleaned of any
foreign materials that could
affect a sealed test air space.
Please refer to the period of
time as specified by the
standard being followed for
the duration of test.
Place the probe in to RH
Hood and cover with Dust
Cap if necessary.
Connect Tramex CMEX II,
WME or MRH III or to take
readings
A probe calibration check is
recommended within 30 days
before the test.
Calibration
Salt
Floor Slab
Floor Slab
Floor Slab
Floor Slab
Dust Cap
Seal the hood to
surface with butyl
tape provided
Sealed test
air space
Hygro-i probe
RH Hood