Technical Specifications 

 

CME5    

Tramex CME5 Concrete Moisture Encounter 

Measuring range 

 

Moisture Content for Concrete: 

0 to 6% 

 

Carbide Method moisture content equivalent for Concrete: 

0 to 4 

 

Carbide Method moisture content equivalent for Anhydrite: 

0 to 1.9 

Display 

Analogue 

Depth of penetration 

20 mm approx 

Construction 

ABS Body 

Power 

2 x AALR6 ALKALINE (included) 

Size 

  

160mm x 85mm x 38mm   

Weight 

344g 

 

Packing List 

 

Tramex CME5 Concrete Moisture Encounter 5 
Hard Carry Case 
Batteries 
Wrist Strap 
Instruction Manual 

 

 

 

 

 

 

 

 

 

 

 

How it works 

 

The Tramex Concrete  Moisture Encounter detects and evaluates 
the  moisture  conditions  within  the  slab  or  screed  by  non-
destructively measuring the electrical impedance. 

 

A  low  frequency  electronic  signal  is  transmitted  into  the  material 
under  test  via  the  electrodes  in  the  base  of  the  instrument.  The 
strength of this signal varies in proportion to the amount of moisture 
present in the material.  

 

The Concrete Moisture  Encounter determines the strength of the 
current  and  converts  this  to  a  moisture  content  value  or  CM 
Equivalent  value  for  concrete  slabs  and  other  cementitious  floor 
screeds, displayed on a large clear analogue dial.

 

 

 

Operating Procedure 

 

To  carry  out  tests  simply  brush  any  dust  from  a  smooth  area  of  concrete  and  also  from  the  electrodes,  switch  on  the 
Concrete Moisture Encounter and place it firmly onto the surface, fully compressing the spring-loaded, signal-enhancing 
contacts  on  the  base.  Read  the  moisture  content  from  the  clear  analogue  dial.  The  Concrete  Moisture  Encounter  is 
calibrated  to  give  percentage  moisture  content  readings  on  clean,  bare  dust  free  concrete  floor  slabs,  with  a  depth  of 
penetration of approx. 20 mm. Readings taken on concrete slabs through paint, coatings, adhesives or other materials on 
the surface should be regarded as qualitative or comparative and not quantitative. 

 