User Manual
TRAQC-8
TRAQC-8 04/2017 EN Rev.4
page : 26
3.3.5
Voltage Output
The Voltage Output option is identical to the “pressure Xmtr” mode, with exception of the
output setting (as set in the next step:” Voltage output Settings”) : Output setting will not be in
pressure units, but in Volt., with fixed Volt steps to go from minimum to maximum. Voltage
output will switch on / off in this menu.
3.3.6
Voltage Output Settings
As explained in “
Xmtr mode settings
”
there are no units to
select: The user is requested to enter Lowest value, highest value
and step value direct in V. Values in excess of the standard range
of 0-11 V will be refused: when ENTERED the instrument will
refuse and display: “ Invalid entry”. The display will
wait for a
valid entry before proceeding to next entry
Following functions ar
e only available when “Current output” or “Voltage
output” has been switched to “on”.
3.3.7
Auto Span Check Mode
Select “On”, automatically the “step time entry” panel is
disp
layed: Enter the required step time using the ‘ARROW’ keys,
then press ‘ENTER’
key (F1).
Select “off ”: the function has been disabled.
3.3.8
Ramp Mode
Select “On”, automatically the “Number of steps entry” panel is
displayed: Enter the required number of s
teps using the ‘ARROW’
keys, and then press ‘ENTER’
key (F1).
Automatically the “Step time entry” panel is displayed: Enter the
required step time in seconds, using the ‘ARROW’ keys.
Press ‘ENTER’
key (F1) to confirm.
Select “off”: the function has
been disabled.
3.4
Electrical input B setup (optional)
This option is under development
3.5
Electrical output B setup (optional)
This option is under development