2.8.

 

Storage     

Prior to transport or contact with the metal block ensure that it has cooled down to a 
temperature below 50°C otherwise there is a risk of severe damage at the instrument. 
 
In order to cool down the TRAQC-3 quickly, set the set temperature to a low 
temperature, e.g. room temperature. Do not switch the instrument of or disconnect the 
mains lead until the instrument has reached a temperature below 50°C 
 

2.9.

 

Recalibration    

The TRAQC-3 is adjusted and tested with measuring equipment in accordance with 
recognized national standards prior to delivery. 
The TRAQC-3 should, depending on the application situation, be inspected at appropriate 
intervals. We recommend you to return the calibrator to Tradinco at intervals  
of max. 12 months for recalibration and service. 
 

2.10.

 

Optional accessories 

 

Carrying case: 

 
A rugged carrying case which offers maximum 

protection to the TRAQC-3 during 
transportation. This case is watertight and 
is secured by two safety latches. It stores 
the TRAQC-3, 3 pockets to store inserts 
and a pocket for the power cord.  

 
 
 
 
 

Inserts: 

 
The following inserts are 

available: 

 
Undrilled 
 
Drilled: 
3.5 mm 
6.5 mm 
4.5 mm + 4.5 mm 
3.5 mm + 4.5 mm 
 
Custom drilled 
 

 

 

Содержание TRAQC-3 FM

Страница 1: ...TRAQC 3 FM Temperature Calibrator User Manual...

Страница 2: ...3 Instrument Label 4 1 4 Scope of this manual 4 2 INTRODUCTION 5 2 1 Heating performance 5 2 2 Control Performance 5 2 3 Functions and connections 6 2 4 Operating instructions 7 2 5 Temperature switc...

Страница 3: ...charge for parts or labor or will provide a replacement in exchange for the defective product In order to obtain service under this warranty Customer must notify TRADINCO INSTRUMENTS of the defect be...

Страница 4: ...ock Attention signal or remark 1 3 Instrument Label 1 4 Scope of this manual This manual contains information for instruments built since September 2008 This manual is a users manual for operating the...

Страница 5: ...ith care and only in areas without hazardous fumes or fire risks and get acquainted with all the facilities offered by this field temperature calibrator 2 1 Heating performance The heating block is ma...

Страница 6: ...nctions and connections 6 7 8 9 10 11 3 4 5 1 2 1 Heat indicator 2 Up and down push buttons 3 Hole for inserts 4 Switching indicator Led 5 Temp switch terminals 4mm 6 Voltage selector 7 ON OF Switch 8...

Страница 7: ...ly XX a code number for manufacturers use only Hr XX The amount of hours the unist has been powered the six digits after the Hr represent the hours up to 99999 9 hrs XXXX see above 4 Place unit under...

Страница 8: ...he mass of the probe may influence the measurement considerably Note If the manufacturer has enabled the most left button one has access to a part of a menu structure Use of this feature is not requir...

Страница 9: ...ensions w x d x h 149 x 159 x 74 mm All specifications at 20 C 2 C 2 7 Maintenance Allow the TRAQC 3 to cool down as described in chapter 2 4 Switch the TRAQC 3 off and disconnect the mains plug Clean...

Страница 10: ...ed and tested with measuring equipment in accordance with recognized national standards prior to delivery The TRAQC 3 should depending on the application situation be inspected at appropriate interval...

Страница 11: ...User Manual TRAQC 3 FM TRAQC 3 LH 10 2014 EN Rev 2 page 11 Tradinco Instrumenten Apparaten b v Radonstraat 250 2718 TB Zoetermeer The Netherlands 31 79 2033133 www tradinco com Email info tradinco com...

Отзывы: