MS 87/./. R 03/03
8
CALIBRATION STEPS
Procedure to access to the calibration menus
DISPLAY ACTION
COMMENT.
1
425
Display before calibration
Hook with no load
2
425
S E
(5 sec)
Press buttons S & E simultaneously for 5 seconds to
access the menu of calibration.
3
CODE
(auto)
(Automatic display)
Access code
4
4567
×
Ø
Press buttons arrows UP & Down to enter code 4567
5
4567
E
Confirm the code by pressing E
6
SEt.4
(auto)
Set-up N° 4.
BASIC CALIBRATION MENU
7
nUL
(auto)
Parameter to set the zero point. (E.g. 750 Hz = 0 Lbs.)
8
nUL
E
Press E
9
425
(auto)
The display doesn’t indicate any incorrect value.
10
425
E
Press E a second time to confirm the zero value.
11
nUL
The setting of the zero point is completed. Press E again
and the display will now show 0. Return the display to
the nuL by pressing E again.
12
nUL
×
Press the UP arrows to pass the next stage.
13
SCAL
(auto )
Define the difference in signal between the zero point
and the working load limit (CALI)
Starting value : 7000 Hz ( Will be automatically adjusted
during the calibration process.)
14
SCAL
E
Press E to view this parameter .
15
7OOO
×
Ø
Use the arrows to correct the value to 7000 (Hz)
If necessary.
16
SCAL
E
Press E to confirm.
17
SCAL
×
Press the UP arrows to pass the next stage.