background image

  Foot Candles

 range/

Fast

 response.

  “Slow” (blue letters)--

  Foot Candles

 range/

Slow

 response. 

2. Turn the meter on by sliding the

  POWER OFF/RANGE switch from the

  OFF position to the desired range. 

To ensure the most accurate reading, 

make certain that the meter is set to the 

correct range:
If the light source is less than 2000 Lux or 

200 Fc, slide the POWER OFF/RANGE 

switch to the 0-1999 Lux/ 0-199.9 Fc 

position.
If the light source is more than 2000 Lux 

or 200 Fc, but less than 20,000 Lux or 

2000 Fc, slide the switch to the 2000-

19990 Lux/200-1999 Fc position.

If the light source is over 20,000 Lux or 

2000 Fc, slide the switch to the 20,000-

50,000 Lux/2000-5000 Fc position. 

3. Place the light sensor directly under

  the light source to be measured.  Hold

  for at least 0.4 seconds. The meter will

  measure the light and display the value. 

4. To hold a measurement on the display,

  slide the DATA HOLD switch to the

  right while a measurement is being

  taken. To cancel the data hold feature,

  simply slide the DATA HOLD switch to

  the left. 

5. Once measurements are completed,

  slide the POWER OFF/RANGE switch

  to the “OFF” position to preserve

  battery life. 

www.

GlobalTestSupply

.com

Find Quality Products Online at:

[email protected]

Содержание 3251

Страница 1: ...acy Lux 5 reading 10 5 reading 100 5 reading 1000 Foot Candles 5 reading 1 5 reading 10 5 reading 100 OPERATION 1 Choose LUX or Foot Candles and Slow or Fast Response by sliding the LUX FC switch and RESPONSE TIME switch to the desired range and response time Fast black letters Lux range Fast response Slow black letters Lux range Slow response Fast blue letters www GlobalTestSupply com Find Qualit...

Страница 2: ...ch to the 2000 19990 Lux 200 1999 Fc position If the light source is over 20 000 Lux or 2000 Fc slide the switch to the 20 000 50 000 Lux 2000 5000 Fc position 3 Place the light sensor directly under the light source to be measured Hold for at least 0 4 seconds The meter will measure the light and display the value 4 To hold a measurement on the display slide the DATA HOLD switch to the right whil...

Страница 3: ...back of the unit away from the unit Remove the old battery and replace it with a new 9 Volt alkaline battery Properly connect the battery Replace the battery cover Incorrectly installed batteries may dam age electronics ALL OPERATIONAL DIFFICULTIES If this unit does not function properly for any reason replace the batteries with new high quality batteries see Battery Replacement section Low batter...

Страница 4: ...NS Traceable is a registered trademark of Control Company CONTROL COMPANY Control Company is ISO 9001 2008 Quality Certified by DNV and ISO IEC 17025 2005 accredited as a Calibration Laboratory by A2LA www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com ...

Отзывы: