STANDARD CALIBRATION
MODE: CALIBRATION
STEP 1:
WHITE VALUES
K 0.08
C 0.07
M 0.08
Y 0.09
EDI
T
4
6
7
5
8
1
2
3
Example: Ihara Calibration Reference Card with Calibration Steps.
Entering Calibration Values
You must match the values displayed on the LCD with
those on the Calibration Reference Card.
If values
displayed differ from those on the Calibration Reference
Card, perform the following procedure:
1.
Press the appropriate
arrow key to scroll to the
2
value you want to change.
2.
Once that value is
highlighted, press the
EDIT key to select.
3.
The value selected will
begin flashing. You can
use the UP ARROW or
DOWN ARROW function
keys to increase or
decrease the value.
IH
A
RA
CALIBRATION REFERENCE
WHITE
CYAN
RESPONSE: STATUS T
Serial No. 70048
DATE: 1997/06/04
MAGENTA
BLACK
YELLOW
STEP 1:
WHITE
K 0.08
C 0.07
M 0.08
Y 0.09
STEP 2:
BLACK
K 1.71
C 1.69
M 1.73
Y 1.79
STEP 3:
SOLID
C 1.57
M 1.57
Y 0.99
?
H
EL
P
M
EN
U
EXI
T
Содержание OMAS
Страница 1: ...OMAS Densitometer COLOR REFLECTION DENSITOMETER OPERATION MANUAL Model R710 R720 R730...
Страница 7: ...7 FEATURES 8 6 7 Bottom View 6 Shoe lock 7 Battery compartment 8 Interchangeable aperture attachment...
Страница 27: ...27 CALIBRATION PROCEDURES STANDARD CALIBRATION QUICK CALIBRATION...
Страница 61: ...61 PRINT CONTRAST Setup Parameters Auto Detection See page 52 Paper See page 53 Solid See page 53...
Страница 85: ...APPENDIX 85 SPECIFICATIONS APPLICATION NOTES...