background image

www.trace2o.com  

19

 

Metalyser

®

 Soils HM4000 

Instruction Manual v3.0

 

 

 

 

3.3.2. PLATING

 

 

This is a process used to form a very thin plate on the surface of the Working Electrode 
and only takes a few minutes to complete. The plating step is necessary prior to the 
analysis of each element, however it is possible to analyse elements consecutively if they 
share a common plating solution (e.g. Cd, Pb and Zn all use HG500 Hg Plating Solution). 
Before carrying out the plating step, the Working Electrode surface must be polished to 
remove any scratches and imperfections (see page 12). 

There are two preliminary steps required for any analysis with the Metalyser

®

. The 

plating step uses the HG500 Hg Plating Solution, HG1000 Thick Hg Plating Solution or 
AU500 Au Plating Solution, added to the SAB, to form a plate on the Working 
Electrode. After plating, the mirror-finish black tip is covered with a plate that is either 
grey (Hg Plating Solutions) or yellow/gold (Au Plating Solutions). 

The element being tested determines the plating solution required: 
HG500 Hg Plating Solution 

– Cd, Pb, Cu, Zn 

AU500 Au Plating Solution 

– Hg and As 

WE1 is used for Cd, Pb, Cu and Zn analysis 
We2 is used for Hg and As analysis 

It is necessary to perform the plating and conditioning before an analysis session or if 
the sensitivity of the analysis decrease. This is because, over time, the plate will reduce 
in thickness and eventually come off. Typically this will be after 20+ consecutive 
samples or a period of 2-3 hours. It is therefore suggested to regularly re-plate the 
electrode; if analysing a lot of samples, it is recommended to plate in the morning and 
again at the beginning of the afternoon session. 

If in doubt, remove the old plate by wiping the surface, polishing the tip of the Working 
electrode to a mirror-finish (with the polishing kit provided), then rinse with the wash 
bottle and perform a new plating. 

When analysing for elements requiring a plate different to the one currently being used, 
the electrode not in use will need to be removed from the sonde to prevent 
interference. It is advised to group together analyses using the same plate to save 
switching between them. 

The conditioning of the electrode occurs when the element of interest is selected in the 

Test Methods

 menu and then choosing condition electrode, which is at the top of the 

list for the selected electrode. The conditioning step should be carried out in sample 
water, and undertakes a run which sensitises the Working Electrode. 

 

 

Содержание METALYSER HM4000

Страница 1: ...METALYSER SOILS HM4000 Portable Heavy Metals Analyser for Soils...

Страница 2: ...lyser 5 3 2 Getting Started 15 3 3 Operating Procedure 18 3 4 Troubleshooting 25 3 5 Interference Effects 27 3 6 Specifications 27 4 Soil Extraction Kit 28 4 1 Getting to Know Your Soils Kit 28 4 2 Sa...

Страница 3: ...t person not trained in voltammetry No complex chemical techniques need to be employed during the course of the analysis The HM4000 is very robust and has been designed to be virtually maintenance fre...

Страница 4: ...SB AC adapters Waste Beakers Sample Analysis Beaker Buffer Kit Instructions Gloves Accessories Cleaning Kit Filtration kit Deionised water Balance Digestion beaker Digestants Clamp Sample Collection B...

Страница 5: ...www trace2o com 5 Metalyser Soils HM4000 Instruction Manual v3 0 3 METALYSER 3 1 GETTING TO KNOW YOUR METALYSER...

Страница 6: ...ing use of the instrument under poor lighting conditions ON OFF Button The ON OFF is used to turn the instrument on and off there is also an optional ten minute auto shutdown feature should you leave...

Страница 7: ...the screen these symbols will appear periodically to provide information about the instrument s status i i Battery fast charge indicator Battery trickle charge indicator Battery fault overheat indica...

Страница 8: ...e it is fully connected To remove the electrodes pull back the connector shroud with the arrow on and the connector will pull off When attaching or removing the electrodes be careful to avoid touching...

Страница 9: ...replacement Note The sonde cable may be fixed to the lid of the Metalyser box using the box closure tabs and the supplied cable clamp Alternatively the user may attach the cable to a suitable fixture...

Страница 10: ...ce after use Always place a few drops of deionised water in the cap before replacing it on the electrode Working Electrode the electrode surface should be examined for cracks and imperfections that wi...

Страница 11: ...ce Electrode Fill Solution to operate Ensure the Reference Electrode is held vertically with the connector upwards and determine that there is enough solution in the Reference Electrode Fill Solution...

Страница 12: ...ting To polish the electrode firstly make sure the glass platen and holder are clean and free of dust or dirt which may cause scratching Place a clean cloth on the platen and dampen the cloth with the...

Страница 13: ...he shaft Sample Analysis Beaker SAB the SAB has a fixed volume so that when full and removed from the water course excess sample will empty out of the holes levelling off at a constant volume Instrume...

Страница 14: ...Addition Used to analyse samples and calculate the unknown concentration by the addition of a known standard Calibrate Used with analyse sample Calibration must be performed before analyse sample is u...

Страница 15: ...fast rate taking approximately 7 hours to fully charge the battery or a trickle charge rate which should be used for overnight top up charge if the instrument is only used once or twice a day From the...

Страница 16: ...om a list or specific location navigate to USB memory drive Windows should install the drivers Once complete the wizard will run again Use the same settings as before Once Windows has installed the dr...

Страница 17: ...stem menu select Backlight Highlight required setting and press OK Press cancel to return to main menu Language Selection The Metalyser comes with English French and Spanish languages pre installed To...

Страница 18: ...o achieving reliable results The following table illustrates which electrode to use for each element If WE1 is to be used in the analysis then WE2 is not part of the circuit and has no role in the ana...

Страница 19: ...Zn analysis We2 is used for Hg and As analysis It is necessary to perform the plating and conditioning before an analysis session or if the sensitivity of the analysis decrease This is because over ti...

Страница 20: ...Choose M2 3 Conditioning Half fill SAB with AU500 Au Plating Solution After plating check tip for yellow gold plate Return AU500 Au Plating Solution to bottle These conditioning steps within the metal...

Страница 21: ...ds to be established just prior to analysis and also after a new plate has been applied to the tip of the working electrode Many factors affect the calibration curve and the greater the time the analy...

Страница 22: ...ng analyses can be found in the application notes Wipe polish WEx Pour Plating Solution into the SAB attach sonde Run Plating on the Metalyser Return Plating Solution to the bottle Tip HMx Buffer into...

Страница 23: ...e around 15mg kg then a 20mg kg standard in this case is sufficient If the user expects the analysis to be 100mg kg then for example a 120mg kg standard would be suitable How to use a pipette 1 Place...

Страница 24: ...left on this screen and then pushing the centre of the joystick to accept it Deposition time Concentration range 30 seconds 50mg kg to 500mg kg 60 seconds 10mg kg to 50mg kg 120 seconds 10mg kg Sugges...

Страница 25: ...er reagents when operating a To check this run the stirrer according to the instructions on page 13 2 Working Electrode is the Working Electrode plated correctly a Is WE1 grey in colour or is WE2 yell...

Страница 26: ...measure high limits by a choosing a shorter deposition time i e change default from 60 to 30 seconds when prompted in the Set Deposition Time menu during the course of the run b diluting the sample by...

Страница 27: ...Type mini USB Type B Bulgin micro series Li Ion Internal Battery Output voltage 3 3V Output current 2200mAh Parameter Lower Limit Upper Limit Arsenic III As III 5 500 Copper Cu 5 500 Mercury Hg 5 500...

Страница 28: ...icient to provide enough extract to perform at least 3 tests One extract can be used for the analysis of Cd Pb Cu As Hg To carry out an extraction procedure you will require the mixer The mixer is ver...

Страница 29: ...sive and harmful to eyes and skin Therefore extra care should be taken whilst handling these chemicals Safety glasses and gloves are provided as part of the soils extraction kit These should be worn a...

Страница 30: ...aduated beaker and then fill up with de ionised water to the 50ml mark 7 Pour the full contents of the graduated beaker into the digestion beaker and swirl to mix 8 Decant from beaker to beaker to ens...

Страница 31: ...displays 0 00 g 6 Use the spatula to gently add the appropriate weight of wet soil into the graduated beaker according to the table below until it shows the correct weight of the soil sample 7 Add app...

Страница 32: ...ove the beaker from the stirrer head and let it settle for 5 minutes 14 Connect a syringe filter to a 5mL syringe Attach a short length of tubing to the filter and use the syringe for taking 3 5ml of...

Страница 33: ...er on the balance provided and press Tare or Zero on the balance and wait until it displays 0 00 g 7 Use the spatula to gently add the dried soil into the digestion beaker until it shows the weight of...

Страница 34: ...nge Attach a short length of tubing to the filter and use the syringe for taking 3 5ml of the digested sample 16 Add the 3 5 mL sample to the sample analysis beaker and fill to the mark just below the...

Страница 35: ...www trace2o com 35 Metalyser Soils HM4000 Instruction Manual v3 0 NOTES...

Страница 36: ...Metalyser Soils HM4000 Instruction Manual v3 0 The Technology Centre Wagtech Court Station Road Thatcham Berkshire RG19 4HZ United Kingdom T 44 0 1635 866772 Fax 44 0 1635 873509 E sales trace2o com...

Страница 37: ...www trace2o com 37 Metalyser Soils HM4000 Instruction Manual v3 0...

Отзывы: