7. Press CAL key to set temperature value. Press ENTER key to confirm; the display shows
“YES/HOLD” indicating that the auto-hold function is enabled.
If the auto-hold function is enabled, the unit will automatically sense a stable endpoint
reading and freeze the reading in the display. If you disable this function, the unit allows user
to freeze the reading in the display manually by pressing the HOLD key.
8. Press CAL key to enable or disable the auto-hold function. Press ENTER key to confirm; the
display
shows “YES/OFF” indicating that the auto-off function is enabled.
When the auto-off function is enabled, if no key is pressed for 8 minutes, the unit will
automatically turn off to conserve power.
9. Press CAL key to enable or disable the auto-off function. Press ENTER key to confirm; the
display
shows “NO/RST” indicating the current status of the reset function.
WARNING
:
The Reset function will restore the unit back to factory default settings; all calibration values
and selected parameters will be reset.
10. Press CAL key to enable or disable the reset function. Press ENTER key to confirm; the
unit returns to measurement mode.
EXIT THE SETUP MENU:
During the setup mode, to exit the setup menu, press ON/OFF key; the unit will return to
measurement mode immediately.