6. CARE AND MAINTENANCE
Routine cleaning.
To avoid any cross contamination of samples the system should be rinsed
through with de-ionised water. Place the inlet tube into a clean beaker filled
with 50ml of de-ionised water and the outlet tube into a second beaker. Run
the instrument for 5 minutes to flush the tubes.
Once a week it is recommended to run a 0.1M solution of hydrochloric acid
through the pipes.
It is not recommended to use acetone or other polar aprotic solvents to clean
the UV Digester enclosure, as these can damage the unit. Trace2o
recommends the use of mild soap solution for routine cleaning of the
enclosure. The use of improper solvents will void the manufacturer’s
warranty.