9.8 Other faults
Error definition, see chapter “Process Input Data - Error Status” on page 60.
Error resetting, see chapter “Index 0x0066: Failure Autoquit” on page 65.
Bit Malfunction Code Cause
Remedy
0
Intensity error
The device checks the
intensity of the received
laser signal continuously, it
was detected a below-
minimum intensity.
1. Clean measuring system optics
2. Clean reflecting foil
3. Rule out an interruption of the laser beam
If the possibility of soiling or interruption of the
laser signal can be ruled out, the device must be
replaced.
1
Device temperature
The measuring system
temperature is outside the
permissible working
temperature. *
This message is a warning. Suitable measures
must be taken so that the measuring system is
operated within the permissible working
temperature.
2
Hardware error
The device has detected an
internal hardware error.
If the error occurs repeated, the device must be
replaced.
3
Laser diode
switched off
The bit is set, if the laser
diode was switched off over
the IO-Link, or the optional
switching input.
Serves only for information purposes.
4
Warning bit
Intensity
The permissible intensity
value has fallen below. *
This message is only a warning and means that
the measuring system optics, or the reflecting foil
is to be cleaned. However, the device operates
error-freely furthermore.
5
Warning bit
Speed-check
The adjusted maximum
speed is exceeded
(chapter 7.4.4 on page 70)
This message is a warning and means that
possibly corresponding measures must be taken,
so that no system components will be damaged.
6
Warning bit
Plausibility
The plausibility of the
measured value couldn't be
guaranteed any more.
This message is a warning and means that
possibly corresponding measures must be taken,
so that no system components will be damaged.
This bit is set also if the device is switched on the
first time in the cold condition. After approx. 1
minute of operating time, if the internally required
minimum temperature were reached, the bit is
reset automatically. Only after this time the
regular operating should be taken up.
7
Measuring range
warning
The measuring range of
0.1 - 25 m was exceeded or
undercut.
This message is a warning and indicates that
appropriate action may be taken to prevent any
equipment from being damaged.
* Limit value can be changed via TRWinProg, see chapter 4.5 on page 56.