3 |
8
DISPLAY MESSAGES
Fe = measurement on ferrous materials (steel or iron)
NFe = measurement on non-ferrous substrates
Err = handling error
IΙFI = wrong substrate, reading beyond measuring range
9
TECHNICAL DATA
Substrate
Iron or steel
Fe-probe
QuaNix®4500 includes:
Non-magnetic metals, e.g.
aluminum, zinc, copper, brass
NFe-Probe
Measuring range
Fe
0 - 3000 μm or 0.00 - 120 mil
NFe
0 - 2000 μm or 0.00 - 80 mil
Resolution
Range 0 – 999 μm: 1 μm
Range ≥ 1 mm : 0.01 mm
resp.
0.01 mil up to 9.99 mil
0.1 mil from 10 to 99.9 mil
1 mil from 100 to 120 mil
Accuracy
+/- (2 μm + 3%*)
resp.
+/- (0.08 mil + 3 %*)
(*) of reading
Minimum object size
Fe
10 x 10 mm² (0.4"x0.4")
NFe
6x6 mm² (0.24"x0.24")
Minimum curvature
convex
5 mm (0.02")
concave
25 mm (1")
Minimum substrate
thickness
Fe
0.20 mm (8 mil)
NFe
0.05 mm (2 mil)
Temperature range
Storage
-10°C to 60°C (14°F to 140°F)
Operation
0°C to 60°C (32°F to 140°F)
Probe
One-point
Power supply
2 batteries AA alkaline
Dimensions
ca.
100 x 62 x 27 mm
(3.9" x 2.4“ x 1.1“)
Weight
ca.
100 g (3.5 oz) incl. batteries