Calibrating the instrument

The  instrument  is  supplied  factory  calibrated.  It  is  only  necessary  to  recalibrate  the 
instrument periodically in line with your in house calibration scheme. 

Warning: 

Altering values in the calibration will affect the operation of the instrument. Do 

NOT 

enter the calibration process unless the instrument is to be recalibrated, failure to 

do so will result in the instrument displaying zero or incorrect readings. 

The  Digital  Krebs  Viscometer  requires  calibrating  at  three  points.  These  points  are  at 
approximately  12,  50  and  100%  of  full  scale  and  have  been  chosen  to  give  maximum 
accuracy  over  the  whole  range  of  the  instrument.  The  calibration  of  the  instrument  is 
carried  out  in  the  grams  range  and  therefore  all  calibration  oil  values  in 
centiPoise/mPas.S must be converted into grams by the following formula: 

              

  

Where, 

L=load (grams) 

n=viscosity of calibration oil in centiPoise/mPas.S 

p=density of calibration oil in g/ml 

The three calibration points are: 

1003 grams 

495 grams 

121 grams 

Due to variations in the viscosity of calibration oils, it may be found that the calculated 
value in grams differs from the calibration point value displayed on the instrument by a 
few units. 

The calculated values must agree with the calibration point values on the 
instrument display before calibration commences. 

If the calculated value differs from the displayed value, adjust the instrument value by 
depressing either the centre left key (STORE) to decrease, or centre right key (UNITS) 
to increase the displayed value. For guidance the following calibration is performed 
using commercially available oils. 

Actual values will vary slightly between batches. 

Содержание SH1349

Страница 1: ...V2 1 0719 DIGITAL KREBS VISCOMETER 480 SH1349 Operating Manual ...

Страница 2: ... a production environment or in the laboratory Features of the Digital Krebs Viscometer 480 Microprocessor control Digital backlit display showing Viscosity in K U mPas S cP Store number Overange indication Calibration sequence Calibration lock Height sensor that prevents the instrument from being operation whilst the paddle is above the level of the tin 9 reading memory Stored values are held wit...

Страница 3: ...scometer 5 Operating the Digital Krebs Viscometer 5 To remove the rotor paddle 5 4 Calibrating the instrument 6 Oils 7 Calibration technique 7 How to make measurements 8 Clearing the memory 9 Microprocessor warning messages 9 5 Instrument Specifications 10 6 Care and routine maintenance 11 7 Spares 12 8 Troubleshooting 13 Display not lighting 13 No reading 13 ...

Страница 4: ...sources of heat 8 Do NOT place the instrument in a strong magnetic field near large electric motors transformers etc 9 The case of the instrument should only be removed by authorised qualified personnel 10 Ensure the unit is placed on a solid and level surface and bolted down using the mounting holes provided 11 Safety critical component If you need to change the power supply unit you must replace...

Страница 5: ...4 2Getting started Packing list Digital Krebs Viscometer 480 Paddle Operating manual Power cord 500ml paint can to BS1262 1989 ...

Страница 6: ...tal Krebs Viscometer Switch the instrument on Fill the Ø 80 mm tin can provided to two thirds full with the sample material Place the tin can on the instruments turntable Push down the handle to its lower location so the rotor paddle is immersed in the sample The rotor paddle will automatically rotate at 200 RPM Observe the turntable and when it is steady note the reading indicated by the pointer ...

Страница 7: ...refore all calibration oil values in centiPoise mPas S must be converted into grams by the following formula Where L load grams n viscosity of calibration oil in centiPoise mPas S p density of calibration oil in g ml The three calibration points are 1003 grams 495 grams 121 grams Due to variations in the viscosity of calibration oils it may be found that the calculated value in grams differs from ...

Страница 8: ...ontents of a full Calibration oil bottle Carefully empty the contents of each calibration oil bottle into separate CLEAN 500ml paint tins to BS1262 1989 Ensure each paint tin and lid is labelled in order to identify their contents Replace lids Switch on the instrument and plug the calibration key into the rear of the instrument The instrument is now in calibration mode The first calibration point ...

Страница 9: ...fore proceeding The instrument display will now show a new value 495 grams Repeat steps 4 9 with the second oil SH0855 The instrument display will now show a new value 121 grams Repeat steps 4 9 with the third oil SH0718 The instrument will display Remove key and emit a warning tone until it is removed Remove the calibration key The instrument is now in normal read mode Check the accuracy of the c...

Страница 10: ...tored the instrument will sound a warning tone This stored value will over write the last value held in memory Holding the read key will keep the paddle rotating until the key is released This allows a longer test to be performed Clearing the memory Once this operation has been carried out the memory should be cleared Simply depress the two centre keys CLEAR simultaneously The instrument will conf...

Страница 11: ...00 r p m 1 Sample container 500ml standard Dimensions 200 mm x 360 mm x 550 mm W x D x H 7 9 x 14 2 x 21 7 Weight 8kgs 17 6lbs Power consumption 30 watts max Electrical supply 200 250V 100 120VAC switchable Working environment This instrument has been designed for indoor use in a temperature range of 5 C 40 C If the instrument is used outside these values its accuracy cannot be guaranteed Associat...

Страница 12: ...ed oil Check tension of the operating handle Lightly grease the main support post Remove any excessive paint deposits from instrument in general Note Always remove and clean the paddle immediately after use Do not allow paint material to come into contact with the quick release chuck Do not allow solvents to come in contact with the keys or the display window ...

Страница 13: ...12 7Spares Calibration key required for instrument calibration SH7120 ...

Страница 14: ...nt switched on If the display is not working after this then check the internal fuse No reading If the display shows zero when taking readings then check the following Is the paddle rotating If yes then the instrument needs calibration If the paddle is not rotating or continues to show zero after calibration then there is an internal fault ...

Страница 15: ...14 ...

Страница 16: ... material is copyrighted by TQC Sheen and may not be reproduced in whole or part by means of information storage retrieval or reproduction without written permission from TQC Sheen Copyright 2018 TQC Sheen ...

Отзывы: