![TPS smartCHEM-pH Скачать руководство пользователя страница 59](http://html1.mh-extra.com/html/tps/smartchem-ph/smartchem-ph_manual_1146435059.webp)
59
17.3 Temperature Troubleshooting
Symptom
Possible Causes
Remedy
Temperature inaccurate and
cannot be calibrated.
1.
Faulty connector.
2.
Faulty Temperature sensor.
Check the connector and replace if
necessary.
Return Temperature sensor for repair, or
replace sensor.
Displays flashing “
M
” when
Temperature sensor plugged
in.
1.
Faulty instrument socket.
2.
Faulty Temperature sensor.
Return the instrument to the TPS factory for
service.
Return Temperature sensor for repair, or
replace sensor.
Reading does not change in
any sample being measured
Auto Stability Function has
frozen display (notice the
).
Turn Auto Stability Function OFF or re-
start sampling (see section 8).