Page 17
6. The
WP-89
will now allow you to enter the value of the second turbidity
standard solution. For example…
180.3NTU pnt2/3
F2:OK
↑↑↓↓
:set
Press
and
to set the value of the turbidity standard solution.
Press
to calibrate the second calibration point.
Press
to quit and retain the current calibration.
7. Place the probe into a the third turbidity standard solution (eg 1800 NTU).
This must have a Turbidity value higher than the second standard.
IMPORTANT :
Leave at least 25mm clearance from the bottom of the
probe to the bottom of the beaker to avoid interference with the reading.
The reading from the probe is displayed. For example…
1803NTU pnt3/3
F2:grab
When the reading has stabilised, press
to GRAB the turbidity standard
solution reading.
8. The
WP-89
will now allow you to enter the value of the third turbidity
standard solution. For example…
1803NTU pnt3/3
F2:OK
↑↑↓↓
:set
Press
and
to set the value of the turbidity standard solution.
Press
to calibrate the third calibration point.
Press
to quit and retain the current calibration.