Safety Information
1.
IMPORTANT:
Read and understand this manual prior to use.
2.
IMPORTANT:
CO instruments may show high readings for up to
30 minutes if taken from warm air to cold temperatures and vice-versa. For
best results let them "soak" in the ambient conditions prior to use.
3.
WARNING:
Substitution of components may impair operation.
4.
WARNING:
This instrument contains a lithium battery which may
l
eak or explode if improperly used.
DO NOT DISPOSE OF IN A FIRE.
5.
WARNING:
Only service in an area known to be free of combustible
gases.
6.
WARNING:
Instruments are not certified intrinsically safe.
7.
WARNING:
To verify operation prior to each day's use a function test
(bump test) should be performed. If the instrument does not pass this
test, full calibration should be performed.
8.
WARNING:
Keep all openings free from dirt, debris and foreign
objects.
9.
WARNING:
Do not use a damaged or improperly operating instru-
ment. Contact a service representative immediately.
ONLY zero instrument in a gas free environment
2