Turning the DC580 on and Connecting a Probe
To turn the DC580 on press and hold the
key for more than two
seconds.
If a probe is not connected the display will read ‘oPEn’. When this
is displayed only the
and
keys operate.
When a probe is connected the DC580 will auto detect which probe
has been connected and will act accordingly.
Hot wire probe -
Make sure the hotwire probe cover has been put
into the closed position before connecting the probe. When the hot
wire probe is connected a 5 second warm up period will begin and
the display will read ‘H---d’ and display the countdown. After the
countdown the DC580 is ready to take measurements. Slide the
cover down to the open position and perform tests.
Vane Probe -
Connect the vane probe and the DC580 is ready to
take measurements.
After power is turned on and a probe is connected the DC580 is
ready to take measurements. Use the setup modes as needed to
adjust units of measure and duct size.
6
www.
GlobalTestSupply
.com
Find Quality Products Online at: