Note:   

(1)   Please   be   careful   when   threading   the   handpiece   to   the   connector;   do   not   use   

excessive   force.   

(2)   When   using   the   handpiece,   please   install   the   drill   or   chuck   to   simulate   actual   

cutting   pressure.   

4.7   

Three   Way   Syringes:      Hold   down   the   ring   collar   to   insert   an   air/water   syringe;   

release   the   collar   ring   when   inserted   to   hold   the   syringe   tip.      To   release   the   tip,   

depress   the   ring   collar;   remove   the   tip,   then   release   the   collar   ring.      The   

air/water   syringe   must   be   fully   inserted   up   to   the   indication   line.   

   

4.8   

Clean   water   bottle:      The   clean   water   bottle   will   supply   the   water   for   the   high   &   

low   speed   handpieces   and   the   three   way   syringe.      Distilled   water   should   be   used   

to   prevent   handpieces   and   three   way   syringes   clogging   from   hard   water   mineral   

deposit   buildup.      When   adding   water,   please   turn   off   the   net   bottle   switch,   

rotate   the   bottle   clockwise   to   remove   the   bottle,   after   filling   bottle,   to   tighten   

turn   bottle   counter-­‐clock   wise.   

   

Notice:      Before   adding   water,   please   turn   off   the   net   water   switch.   

   

4.9

 

Air   System:      Prior   to   packaging,   the   oil-­‐free   compressor   has   passed   all   quality   

tests.      Open   the   pressure   switch   will   start   the   air   compression.      The   air   

compressor   pressure   gauge   should   slowly   rise   from   0   bar   to   8   bar.      When   the   

pressure   in   thank   drops   below   5   bar,   the   air   compressor   should   start   

automatically   until   the   pressure   build   up   to   8   bar.   

   

Notice:      

Do   not   operate   or   disassemble   this   equipment   without   professional   

guide   in   order   to   avoid   injury   or   unit   damage.   

   

5.

 

Operation   and   Maintenance   

   

5.1

 

Operation:      The   machine   should   operate   in   normal   room   temperature   from   5-­‐

40’C   and   relative   humidity   not   greater   than   80%.      Surrounding   area   of   the   

machine   should   be   clean,   dry   free   of   corrosive   gas,   well   ventilated   with   indirect   

sunlight.   

   

Notice:

   The   machine   should   be   placed   firmly,   evenly   leveled   on   a   hard   flat   

surface.   

   
After   Opening:      Check   to   see   if   the   machine   is   in   good   condition   and   has   all   its   

components   outlined   in   the   packing   list.   

   

Содержание PC2635

Страница 1: ...Portable Dental Unit PC2635 Operational Manual Notice Before Operating the unit please read the manual carefully and reference it for future maintenance procedures...

Страница 2: ...n Any damages to any of the wiring cables or attachments are to be replaced immediately The unit should be kept clean and dry there should not be any noticeable oil leaks on any surface of the machine...

Страница 3: ...w 2 Symbols 3 Operational Overview 4 Main Functions 5 Operation and maintenance 6 Operating Diagram 7 Standard Configuration 8 Technical Specifications 9 Portable Unit Troubleshooting 10 Air Water Cir...

Страница 4: ...s clinics and hospitals for the treatment of oral health Equipped with a self contained air compressor air tank air filter and regulator clean water bottle three way syringe high speed hand pieces tub...

Страница 5: ...e air tank is collected at the bottom of the tank where it can be discharged The pressure regulator controls and maintains the air pressure within the air tank You can adjust the air pressure by raisi...

Страница 6: ...after filling bottle to tighten turn bottle counter clock wise Notice Before adding water please turn off the net water switch 4 9 Air System Prior to packaging the oil free compressor has passed all...

Страница 7: ...erve the pressure gauge on a full tank to see if the reading on the gauge drops If there is no detectable air leak open the air outlet valve next to test the pressure switch When the air pressure drop...

Страница 8: ...enance Routine handpiece maintenance including oil and lubricant should prolong the life of a handpiece 5 7 Handpiece Nozzle Cleaning Weekly handpiece air water nozzle maintenance should prolong the l...

Страница 9: ...Three way syringe 7 Tank 16 High Speed HP tube 23 Handpiece water adjustor 8 Pressed drain valve 17 Power switch 9 Water Bottle 18 Foot pedal 24 Handpiece air adjustor 7 Standard Configuration 4 Hole...

Страница 10: ...of air compressor is too high Touch the shell of the air compressor by hand Cool the air compressor use it until it is cooling 2 Compressor cannot be stared The forward valve fails to work Take off t...

Страница 11: ...sure rise or use a manostat The water in water tank has been used up Check the water volume of the water tank Replace the tank 7 The handpiece cannot spray water while in use Air water distributing va...

Страница 12: ...10 Air Water Circuit Diagram 10 1 Electrical Circuit 10 2 Air and water circuit 11 Air Water Circuit Diagram 1 PC2635 1 pc 2 Product manual 1 pc...

Отзывы: