Markings
Figures
Notice:
Risk of losing accessories and documents
Caution:
Risk of injury or malfunction
Read carefully of the manual.
Ground Protective
Electron garbage bin
Producing date
On
Off
Troubleshooting
Unit will not power on:
Check power cord.
Check power switch.
Check resettable fuse.
If no green light is illuminated then
there is no power to the main switch.
Check fuse. Check outlet.
Suction is Low:
Verify the Saliva Ejector Valve lever is
opened.
Verify there are no kinks in the tubing.
Verify waste container is secured.
Check saliva Ejector screen located
under the rubber tip.
Air Venture Valve clogged. Contact
support for further instructions.
Low handpiece pressure:
Try adjusting the HP pressure using
the adjustments in operation of
controls.
If adjustment knobs do not adjust the
pressure, check both HP positions. If
the result is the same on both
positions the main regulator may
need adjusting. Call support for
further instructions.
Unit is very loud when the
motor is running.
Verify shipping screw have been
removed.
Unit make a pop sound when
the compressor is running:
If you hear a consistent pop sound
while motor is running, discontinue
use immediately. The safety pressure
valve is active. This Valve releases
excessive pressure in the event the
automatic pressure shut off fails.
Contact TPC support for further
instructions.