Laguna 2000 Chair Troubleshooting 

Problem  

Cause  

Solution  

Dental Chair does not move 
up or down.  Chair makes no 
sound.  

No power  

Blown fuse  

Power switch off  

Blown transformer  

Check power source  

Check fuse  

Check power switch 

Check transformer voltage 
12vac 

Dental Chair will not move, 
but makes a beep sound when 
controls are pressed  

Safety switch plate  

Limit switch memory erased  

Check safety plate  

Re-program limit switches. 
Refer to motions of limit 
controls 

Chair makes noise when 
motor is running. Squealing  

Damage to drive extension  

Check drive extension for 
damage 

Dental Chair does not go low 
or high enough  

Limit switch moved or is 
damaged  

Adjust limit switch or replace 
is damaged  

Unable to program pre set 
position 1, 2 and 3  

Press SET to program presets. 
Button may be damaged  

Press Set. Replace set button 
on touchpad.  

My chair is on but there is no 
control out of the touch pad 

Touch pad is not connected  

Touch pad is burnt out  

Connect touch pad 

Replace touch pad 

My head rest does not stay in 
the position that I put it in 

Set screws in backrest are 
loose.  

Tighten set screws 

Chair arm rest moves very 
Easley.  

Screw is loose that holds 
assembly together 

Tighten screw 

 

 

 

 

 

Содержание Laguna 2000

Страница 1: ...1 www tpcdental com www tpcdental com Laguna 2000 Electromechanical Patient Chair Installation Operation User Manual TPC 851 S Lawson St City Of Industry CA 91748 P626 810 4337 Fax 626 810 4245...

Страница 2: ...Table on Contents Page GENERAL INFORMATION 3 Installation and coordination 5 Motion Limit Controls 10 Operation instructions 11 Overview of controls 12 Main PCB Board 14 CLEANING DISINFECTING GUIDE 18...

Страница 3: ...bel is located on the front lower left cover under the seat cushion The dental chair is marked with a Risk Class symbol This label is located on the front of the dental chair base next to the main pow...

Страница 4: ...r forklift to move the dental chair Do not attempt to move stacked dental chairs Only store the dental chair in dry cool place The dental chair is marked with a main electrical grounding point symbol...

Страница 5: ...port arm to the backrest bracket Use the three supplied Allen screws 2 Install chair backrest as shown in the images below Using the 8 Allen screws secure the backrest frame in place 3 Remove the side...

Страница 6: ...ww tpcdental com www tpcdental com 4 Install the chair bracket as shown use the included 4 Allen screws 5 Re install the chair side cover once the bracket is installed 6 Mount the Cup to the chair bra...

Страница 7: ...al com 7 Install the backrest cushion onto the frame Make sure you align the mounting points before sliding into place 8 Install the seat cushion onto the seat frame Make sure you align the mounting p...

Страница 8: ...tpcdental com 9 Once you have confirmed the seat cushion is in place slide it toward the backrest to lock it in place Secure the seat cushion in place using the two supplied Allen screws 10 Insert th...

Страница 9: ...9 www tpcdental com www tpcdental com 11 Install the Pump cover and rear base covers Use the supplied screws to secure the covers in place...

Страница 10: ...e nut is not secured you can adjust the switches position to activate at a different height Once adjusted secure the adjusting nut Emergency Stop Button The dental chair possesses an emergency stop bu...

Страница 11: ...11 www tpcdental com Laguna 2000 Electromechanical Patient Chair Operating instructions...

Страница 12: ...o recline press down on left side of foot control disc 3 Preset Operation Momentarily press pre set switch 1 or 2 on the foot control 4 Automatic Return Momentarily press 0 switch on foot control to a...

Страница 13: ...rneath 7 Armrest Rotation Either armrest can be rotated upward by grasping the end portion of armrest pulling upward 8 Headrest Adjustments Height of headrest can be adjusted simply by pulling headres...

Страница 14: ...14 www tpcdental com...

Страница 15: ...15 www tpcdental com...

Страница 16: ...16 www tpcdental com...

Страница 17: ...17 www tpcdental com...

Страница 18: ...sinfectant to soak in or dry on the surface wipe off with soap and water PLEASE KEEP LIQUIDS AWAY FROM ELECTRICAL CONNECTIONS Most electrical control switches ARE NOT hermetically sealed When a switch...

Страница 19: ...ms against the freight carrier must be initiated at the time the damaged items are received The claim is the responsibility of the customer We are improving our products on a continuous basis We reser...

Страница 20: ...when motor is running Squealing Damage to drive extension Check drive extension for damage Dental Chair does not go low or high enough Limit switch moved or is damaged Adjust limit switch or replace i...

Страница 21: ...of 600 lbs or damage may occur Don t let more than one patient sit in the chair at one time Don t stand anywhere on the dental chair Don t operate the dental chair around any flammable liquids Don t...

Страница 22: ...e screws shown 2 Remove the front cover 3 Remove the bracket cup from the chair bracket 4 Install bracket onto the dental chair 5 Install the side cover over the bracket as shown Repalce screws 7 Re i...

Страница 23: ...ight head to arm Switch on off Adjustment To switch the lamp On and Off press and release the switch to the LEFT or RIGHT side When the lamp is switched On Off the device will beep once DON T push and...

Страница 24: ...1 www tpcdental com Mirage Dental Delivery Unit Model 2000 2015 Installation Instructions TPC 851 S Lawson St City Of Industry CA 91748 P626 810 4337 Fax 626 810 4245...

Страница 25: ...Inventory 3 Installation and coordination 4 Optional Cuspidor Installation 10 Cuspidor Plumbing Connections 11 Junction Box Connections 12 Post Mount Utility Center Adjustments 13 Pressure Adjustment...

Страница 26: ...Cosmetic beauty ring 2 Post mount utility center Utility center Junction box frame cover Master controls attached to the umbilical Misc beauty rings and quick connects HEV SE valves with tubing Water...

Страница 27: ...w 3 Leveling Chair bracket The chair bracket has a total of 8 set screws Use the lower set screws to adjust the PMU left right front and back This will make large adjustments to the tilt of the post t...

Страница 28: ...verify if the PMU is aligned straight with dental chair armrest Then proceed to tighten the set screws 5 Verify the PMU is level before tightening the set screws on chair bracket There are 8 set screw...

Страница 29: ...adjust its angle 7 Remove the PMU side cover To remove it pull on the lower lip to detach its magnetic connection Place it out of the way to ensure it does not get damaged This side cover will remain...

Страница 30: ...e set screw to secure the light post in place 9 Take the delivery unit flex arm out and place the additional cosmetic ring on it as shown below Once in place take all lines and feed it through the bus...

Страница 31: ...ou fed through in the previous step Take the touch pad wire shown below and feed it through the chair bracket as shown 11 Route the touch pad cable along the chair bracket channel Then feed it through...

Страница 32: ...the flex arm Grey female from umbilical to Grey male from the flex arm Purple female from umbilical to Purple male from the flex arm Dark Red female from umbilical to Dark Red male from the flex arm B...

Страница 33: ...he cuspidor is completely seated in the bushing 15 Take the cup fill and bowl rinse spout and insert them into the corresponding sockets When inserting the spouts into the sockets make note of the not...

Страница 34: ...p 17 Remove the cap from the large water supply line Connect the water supply line to the cuspidor Remove the cap from the signal air line then connect it to the cuspidor 18 The 2000C cuspidor possess...

Страница 35: ...ge blue line that exits the back of the master control valve 21 Once the connections are made you may open the angle stops or equivalent shut off valve Once the valve is open you may check the mater c...

Страница 36: ...to supply the bottle Tension Adjustments 2 In a situation that you need to add tension to the delivery unit flex arms please see the following locations NOTE tension adjustments are only a temporary f...

Страница 37: ...arm See image Once the screws are loose use the two adjustment screws in the center on each side to level the head To adjust the unit head tilt remove the flex arm end cap closest to the delivery uni...

Страница 38: ...rom umbilical to delivery unit arm 1 Yellow 6 X 4 Supply line T to unit head 2 Black 6 X 4 from output from foot control 3 Small Yellow Small Yellow Master Switch return 4 Small Gray Small Gray supply...

Отзывы: