Toshiba LF430/LF620 Скачать руководство пользователя страница 8

TIC-LF434H

 

 

 

Separate type LF430/LF622 flowmeter and LF434/LF622F flowmeters 

Grounding with 100

Ω

 or less  

ground resistance 

Instrument panel : Ordered separately 

I/O cable

(CVV-S)

Grounding with 100

Ω

 or less 

ground resistance 

wire 5.5mm

2

 or more 

Current output (4

20mAdc) 

or PROFIBUS 

Power supply 

Power cable

(CVV) 

Digital input 

(20

30Vdc)

 

Signal common for DI and DO

 

Digital output 1

 

Digital output 2

 

Grounding with 100

Ω

 or less 

ground resistance 

or Modbus

 

Signal cable 
(2-wire shielded hard-rubber sheathed cable)

Connected detector

Excitation cable 
(3-wire shielded hard-rubber cable)

Thick walled steel conduit 

Power switch 

(External double-pole power switch) 

 

Figure 7.  Separate type LF430/LF622 and LF434/LF622F flowmeters wiring Diagram 

 

Table 2. LF620, LF620F,  LF622 and LF622F Converters Signal Table 

 

Symbol Description

Cable

L1 (

Power supply 

Power cable (CVV) 

L2 (

GND 

Ground (for arrester)

FG Frame 

ground

DI 

Digital Input (20

30Vdc)

I/O cable (CVV-S) 

DO1 

Digital Output 1

DO2 

Digital Output 2

COM 

Signal Common for DI, DO1, DO2

 

 

Current Output (4

20mAdc) 

or PROFIBUS 

 

Shielded cable for 

PROFIBUS-PA 

 

X

Excitation cable 

 (for LF622,LF622F only) 

Y

Excitation Output

E

A

Signal cable 

(for LF622,LF622F only) 

B

Signal Input

G

T+ )

Twisted-pair polyethylene 

insulated vinyl sheath cable 

(JKEV,AWG24(0.2mm

2

))

 

T- Modbus(-)

TG Modbus(GND)

Note: 

Symbol of the terminal is changed as follows for Modbus.

 

                                                       DO2 

 T+,   DI 

 T-,   COM 

 TG 

 
 
 
 

Содержание LF430/LF620

Страница 1: ...communications protocol for factory and process automation that the PROFIBUS Organization recommends Instead of analog control with a conventional analog signal 4 20 mA it is fieldbus which digitizes all signals Flowmeters support PROFIBUS PA Modbus is the communication protocol that Modicon Inc developed Physical layer is RS485 Figure1 Configuration LF430 LF620 LF430 LF622 LF434 LF620F LF434 LF62...

Страница 2: ...ange is required to maintain the specification performance Note The test results were obtained and demonstrated at TOSHIBA s flow calibration facility Fuchu Japan Drinking water approvals Hard rubber liner NSF Note Hard rubber liner is available from 4 to 24 Model LF430 and LF434 Detectors Mounting style Flange connection type Fluid pressure 15 psi or 1 0 bar 0 1 MPa to the nominal pressure of the...

Страница 3: ...put non polarity Number of outputs One point Output capacity 150Vdc 150mA maximum or 150Vac peak to peak 100mA maximum Note DO2 cannot be used with the Modbus communication DO1 and DO2 functions One of the following functions can be assigned to DO1 std and or DO2 opt Pulse output available only for DO1 DO2 Pulse rate Max 10kHz 10 000pps DO1 Max 100Hz 100pps DO2 Over 1kpps auto setting Pulse width ...

Страница 4: ... to 36Vdc 110Vdc allowable voltage 90 to 130Vdc Surge protection Arresters are installed in the power supply and a current signal output circuit to help protect the meter from lightning and improve personnel safety Case Aluminum alloy equal to IP 67 Coating Acrylic resin baked coating pearl gray colored Cable connection port Cable glands LF620 and LF622 without cFMus Approval Provided as standard ...

Страница 5: ... 140 205 252 4 approx 5 0 25 160 208 265 4 approx 7 5 40 170 216 291 4 approx 9 5 50 180 225 308 4 approx 11 5 80 230 239 339 8 approx 16 5 100 240 257 367 8 approx 20 5 150 260 287 430 8 approx 36 0 200 300 313 483 12 approx 49 0 250 350 336 539 12 approx 105 300 400 363 593 12 approx 113 350 450 380 640 16 approx 130 400 500 401 684 16 approx 173 JIS 10K dimensions Meter size mm L1 mm L2 mm L3 m...

Страница 6: ...ight kg 15 140 147 194 4 approx 4 0 25 160 149 207 4 approx 6 0 40 170 158 233 4 approx 9 0 50 180 167 250 4 approx 11 5 80 230 181 281 8 approx 17 5 100 240 199 309 8 approx 22 0 150 260 229 372 8 approx 37 0 200 300 255 425 12 approx 52 0 250 350 278 481 12 approx 108 300 400 305 535 12 approx 121 350 450 322 582 16 approx 145 400 500 343 633 16 approx 188 450 550 372 707 20 approx 208 JIS 10K d...

Страница 7: ...OTE In the case of the integral type the signal line A B G and the excitation line E X Y are the connector joints Grounding with 100Ω or less ground resistance 1 Locate an external double pole power switch on the power line near the flowmeter within easy reach of operation Use the appropriate switch rating as shown below Switch rating 250Vac 6A or more In rush current 15A or more Figure 6 Combined...

Страница 8: ...ed steel conduit Power switch External double pole power switch Figure 7 Separate type LF430 LF622 and LF434 LF622F flowmeters wiring Diagram Table 2 LF620 LF620F LF622 and LF622F Converters Signal Table Symbol Description Cable L1 Power supply Power cable CVV L2 GND Ground for arrester FG Frame ground DI Digital Input 20 30Vdc I O cable CVV S DO1 Digital Output 1 DO2 Digital Output 2 COM Signal C...

Страница 9: ...ing nicked or damaged In addition for cables be careful of allowable maximum bend diameter Do not allow excessive twisting or bending of cables 5 Consider installing a PROFIBUS PA arrester in the communication path of PROFBUS PA so that the electromagnetic flowmeter will not be affected by lightning etc 6 The electromagnetic flowmeter is not equipped with terminating resistors Use the terminating ...

Страница 10: ...354 450 18 2 500 4 366 20 000 26 029 500 20 3 000 4 244 25 000 26 354 600 24 4 000 3 930 40 000 29 283 Note The unit of gal min is not exchanged converted by m3 h About establishment environment Do not store or install the flowmeter Where there is direct sunlight Where excessive vibration or mechanical shock occurs Where high temperature or high humidity conditions exist Where corrosive atmosphere...

Страница 11: ...0K Z Other Electrode Material Note 5 B 316L stainless steel C Ti titanium D Pt Ir platinum iridium E Ta tantalum F Hastelloy C Equivalent Z Other Lining Materials Note 5 C Teflon PFA 16 Note 8 Note 8 D E EPDM Rubber 3 18 Hard Rubber with NSF approval 4 24 Grounding Ring Material Note 5 C 316 stainless steel D 316L stainless steel E Ti titanium F Ta tantalum G Pt Ir platinum iridium H Hastelloy C E...

Страница 12: ...l SUS304 Accessory for pipe installation BNP material SUS304 2 Digital input output Digital output points 2 DO1 DO2 Digital input point 1 DI 1 2 3 Current output and Communication function Current output HART communication PROFIBUS communication Current output is not functional Current output Modbus RS485 communication Digital outputs 2 DO1 DO2 and Digital 1 DI input are not functional 1 2 3 Power...

Отзывы: