background image

LQ500  

EJL-106J 

prepare a loose mechanism in advance. 

(9) To minimize the impact of the bubbles, it is 

recommended that the meter be installed on a 
location as far as possible from the pipe outlet for 
air release but still within the distance where a 
reasonable degree of hydraulic pressure is 
applied. 

(10) In the event that the density meter may no 

longer be full of the fluid while the pump is shut 
down or the density distribution in the density 
meter may become uneven, make sure to take 
measurements only while the pump is operating 
by using the external synchronized function. 

(11) Take necessary measures to prevent vibration 

from a pump or other equipment applied to the 
density meter transmitted through the piping. 

(12) On both the upstream and downstream sides of 

the density meter, install isolation valves. 
Furthermore, between these valves and the 
density meter, install the sampling port, the zero 
water supply port, the air release port, the drain 
port with a stop valve attached respectively. In the 
event that the flow of the pipe line cannot be 
stopped, provide a bypass pipe halfway with a 
stop valve attached. When performing zero point 
calibration, these are needed to discharge the 
measured matter out of the density meter through 
its drain port and fill up the meter with fresh water 
of zero density. (See Figure 6 and Figure 7) 

(13) As for gaskets to be used in piping, select the 

one with the dimension conforming to the flange 
standard and of the material appropriate for the 
substance to be measured. 

(14) The front side of the density meter's converter 

section is equipped with a density display section. 
When installing the meter, choose a location and 
direction in which this density display section 
will be easily visible. 

(15) If the cover of the density meter is removed or 

the density meter is disassembled while the meter 
is powered, radio waves will leak out. (However, 
the amount is about equal to PHS and one tenth of 
mobile phones.) 

(16) For both horizontal and vertical piping systems, 

install a support stand under the density meter 
with bolts put through the installation holes (M8 
size) on the bottom of the density meter. Then 
install the support stand together with the density 
meter to a solid ground or wall. Bolts (four M8 
bolts) are not attached as accessories of the 
density meter. Please prepare these bolts 
separately 

NOTE: 

 Zero point water valve: 

Used to supply drinking water (density or consistency 
0%) to the detector pipe for zero point adjustment. 
Install this valve at the top of the pipe in the case of 
horizontal installation. It is recommended that a 1-inch 
ball valve be installed on the top of the pipe and zero 
point water supplied through this inlet using a vinyl 
hose etc.  

 

Note: If valve water pipe is connected to this valve, air 

cannot be extracted. Therefore, another valve (vent 
valve) is needed to extract air. 

 Vent valve:

 

Used to vent process fluids to open air when 
performing zero adjustment. This helps the drinking 
water (density or consistency 0%) enter the detector 
pipe easily. Install this valve on the top of the pipe in 
the case of horizontal installation. 

 Drain valve: 

Used to drain the fluids before supplying drinking 
water (density or consistency 0%) to the detector pipe 
for zero adjustment. Install this valve at the lowest 
point of the pipe. It is recommended that a 1-inch ball 
valve be installed at the lowest point of the pipe. 

 Sampling valve: 

Used to extract fluids for manual analysis. Install this 
valve to the side of the pipe in the case of horizontal 
installation. It is recommended that a 1-inch ball 
valve be installed to the side of the pipe. 

 

 

Stop   
valve 

Bypass

pipe

Zero point water valve 

Flow direction 

(from down to up directions) 

Drain pipe 

Vent valve 

Drain valve 

Stop valve

LQ500 

Stop valve 

Zero point water pipe 

 

Figure 6. Recommended Installation 

(vertical installation) 

 

Figure 7.  Horizontal installation 

(*inside the pipe is always filled) 

LQ500 

 

Содержание Density (Consistency) Meter LQ500

Страница 1: ...ollowing cases Possibility of unfilled condition a When it is installed at the discharge of a pump b When installation is horizontal and unfilled condition occurs inside the pipe c A process where the pipe becomes unfilled when the operation is stopped Cable for power supply Detector converter Ground resistance 100 ohm or less Density consistency measurement output 4 20mA Cable for communication D...

Страница 2: ... reagent Electric resolution Meter size 50 mm 2 80 mm 3 to 300 mm 12 Electric resolution 0 002 TS 0 001 TS Note 1 Above values are the results of computing ability in the phase measurements of the converter Note 2 Density TS consistency determination resolution for sample reagent Meter size 50 mm 2 80 mm 3 to 300 mm 12 Density TS consistency determination repeatability 0 1 TS 0 05 TS The density T...

Страница 3: ...ximum working pressure ANSI Class 150 1 MPa 150 psi DIN10 and BS10 1 MPa 10 bar DIN 16 1 6 MPa 16 bar JIS 10K 1 MPa 10 kgf cm2 Note Each product was passed a hydraulic test under twice pressure rate for 15 minutes toward the specification Fluid temperature 0 to 100 deg C 32 to 212 deg F without freezing and bubbles conditions Allowable fluid conductivity Meter size Fluid conductivity 50 mm 2 20 mS...

Страница 4: ...itable for your CIP or not before cleaning 2 The smoothness inside the pipe on this material is Type Meter size Smoothness Standard type 50 to 200 mm 2 to 8 No buffing 250 300 mm 10 12 Buffing 150 Option type 50 to 300 mm 2 to 12 Buffing 150 3 State the wetted materials when you choose these options 4 The materials of them are changed for abrasive applications Need to choose the specification code...

Страница 5: ...density TS consistency measurement Contact open Stops density TS consistency measurement Density multiplier switching signal In order to achieve selecting up to 4 kinds of liquid concentration measurement independently as maximum Specification Two voltage signals described below are required Input voltage H level 20 to 30 Vdc L level 2 Vdc or less Input resistance Approx 3k ohm Internal conductivi...

Страница 6: ...ation N when the external synchronized operation starts LQ500 measure the phase of microwave and calculate the density TS from the phase changes Because the phase is limited from 0 degree to 360 degrees LQ500 use the rotation N when the phase is over 360degree But the range of density is wide over the 360 degree LQ500 may not know the number of the rotation N when the pump starts If the density is...

Страница 7: ...x 44 97 lb Approx 42 200 8 320 12 6 290 11 4 300 11 8 Approx 50 Approx 52 Approx 54 119 lb Approx 48 250 10 300 11 8 315 12 4 350 13 8 Approx 65 Approx 68 Approx 68 150 lb Approx 64 300 12 360 14 2 340 13 4 350 13 8 Approx 78 Approx 85 Approx 99 218 lb Approx 76 110 4 3 119 4 7 Detector Converter communication cable port Steel pipe 50A 2 360 14 2 390 15 4 234 9 2 Density signal output port Contact...

Страница 8: ...t measure accurately The range of measurable density less than 1 second c 1 000 Meter Size 50 80 100 150 200 250 300 Density TS 16 8 10 5 8 4 5 6 4 2 3 4 2 8 500 19 7 600 23 6 Maintenance space Back Front Detector Converter RF part A 500 mm 19 7 maintenance space is also needed above unit 500 19 7 600 23 6 600 23 6 500 19 7 Maintenance space Unit mm inch Figure 5 Clearance space Piping Precautions...

Страница 9: ...d radio waves will leak out However the amount is about equal to PHS and one tenth of mobile phones 16 For both horizontal and vertical piping systems install a support stand under the density meter with bolts put through the installation holes M8 size on the bottom of the density meter Then install the support stand together with the density meter to a solid ground or wall Bolts four M8 bolts are...

Страница 10: ... Screws at the terminals are needed to tighten with 1 2 N m torque 1 4 N m is maximum 10 Do not turn on the power supply under the uninstalled condition 11 Each cable in the communication cable and power supply cable between detector and converter has banded marks for each terminal Connect them correctly without any mismatches L1 L2 Tx TX Rx Rx SG FG 24V 0V FG Ec Ec FG Tx Tx Rx Rx SG FG 24V 0V FG ...

Страница 11: ...nction is used By connecting the cable lead of the optional BF100 configurator to the density TS consistency measurement output you can operate the LQ500 from remote places 1 Make sure the load resistance of 4 20mAdc line of output is between 240 and 750 ohm and the load capacitance is 0 25µF maximum 2 The BF100 s cable lead can be connected to the LQ500 anywhere along the current output line For ...

Страница 12: ...pe for abrasive application Note 2 Cable length between detector RF part converter A 10 m 32 8 ft standard B 20 m 65 6 ft C 30 m 98 4 ft D 40 m 131 2 ft E 50 m 164 ft X None Note 2 Note 1 The differences between standard type are RTD sensor and Applicator window Note 2 Toshiba recommends to using our specified cable Table 5 Type Specification Code BF100 Hand held Terminal Model Specification Code ...

Отзывы: