![TORREY PINES SCIENTIFIC EchoTherm SC25 Скачать руководство пользователя страница 6](http://html.mh-extra.com/html/torrey-pines-scientific/echotherm-sc25/echotherm-sc25_operating-manual_1134657006.webp)
Torrey Pines Scientific, Inc.
6
VI. GENERAL DESCRIPTION
Torrey Pines Scientific SC25 units are Peltier driven for chilling & heating. They come with
a universal power supply and the chilling/heating module. The units have two moving
parts, the DC fan that cools the unit and the DC motor that oscillates the plate. Everything
else is solid state and should last years without problem. All functions of the unit are
accessible from the front panel via the membrane switch and accompanying digital display.
HEATER/CHILLER PLATE
The surface is a very flat aluminum plate designed for good contact with any flat surfaced
item placed on it. The plate size of the SC25 is 2.875" (73 mm) x 4.375" (111 mm). It will
chill and heat quickly without a load on it. The temperature of the plate is sensed by a
platinum RTD mounted under the plate. The temperature controller in the unit compares
the plate temperature with the target temperature and instructs the Peltier module to heat
or chill the plate as required to keep them equal.
ORBITAL MIXER
The plate, and therefore the sample blocks placed on it, can be instructed via the
membrane switch to rotate from 200 to 1000 rpm. The plate is supported by a flexure
leaf spring mechanism that is driven by a motor at the rear of the unit. The orbit is
about 2mm peak to peak and works well for mixing small samples in centrifuge tubes at
faster speeds and larger samples in larger vessels at slower speeds.
This is a rugged
mechanism, but take care not to drop anything heavy on the plate surface
.
PROGRAMMABLE
All SC25 units enable the user to define up to 5 individual programs consisting of up to 10
steps per sequence. All sequence steps are stored in non-volatile memory.
TIMER
All units have a countdown timer reading in hours, minutes, and seconds. It can be set to a
maximum of 99 hours 59min 59sec. The timer will be displayed at the same time as the
SET POINT and PLATE TEMP. It has a user settable AUTO-OFF as well. This works to
turn the chiller/heater target temperature off when the timer counts to zero.
ALARM
The units have an audible alarm that sounds for one minute when the timer counts down to
zero. Touching the UP ARROW will turn the alarm off during this first minute. However, if
the alarm sounds for the entire minute, it will shut-off the sound automatically. When the