SETTING THE UNIT TO “IDLE MODE”
Setting the IC50 to “Idle Mode” will turn off power to the heater/chiller assembly and
allow the sample to heat up or cool off depending on the ambient conditions. It is
essentially turning the set point off, which is why the set point on the LCD indicates “off”
when the unit is in Idle Mode. Setting Idle Mode is accomplished by entering the Menu
mode, scrolling to “Set Idle Mode”, and pressing both arrow buttons. Clearing Idle
Mode is the same as setting it but the Menu option will read “Clr Idle Mode” when the
unit is in Idle Mode. Additionally, the Idle mode can also be cancelled by pressing either
the Up or Down arrow buttons on the keypad. When Idle mode is cancelled, the set
point for the unit will return to the value that it was when the unit entered Idle Mode.
See Figure 4 and Figure 5 for details on setting and clearing Idle Mode.
Figure 5:
Menu Map to Set Idle Mode
Torrey Pines Scientific, Inc.
12