value, the accumulator switches off, thereby preventing a full discharge
and further failure of the accumulator.
4.2.7. The LCD of the externally powered flowmeter has a backlight. In case of
external power failure and switching to the internal power supply mode,
the display backlight goes off.
4.3. Troubleshooting
4.3.1. List of possible faults, failures and alarm situations detected by the
flowmeter and displayed on the LCD with "
X
" character in the
"
ERR = - - - - - -
" line is given in Table 5. A sign-position code is read
from left to right.
Table 5
Sign position number
Explanation
1
Hardware failure
2
Battery voltage is less than the specified threshold
3
USS not detected
4
Current flowrate value exceeds maximum
5
Flowrate value violates the upper threshold specified for
Universal output N1
6
Flowrate value violates the lower threshold specified for
Universal output N1
4.3.2. The same information can be read via the communication interface from
the
Status
tab of the "Viewer BUF-311" program (see section 4.1.5).
4.3.3. If "
X
" symbol is displayed in position 1, the flowmeter shall be forwarded
to the manufacturer for repair.
If "
X
" symbol is displayed in position 2, battery replacement is re-
quired.
4.3.4. If "
X
" is displayed in other positions and/or in case of no pulses at Uni-
versal output N1, check the following:
- Flowmeter and Secondary Power Source input voltage is present and
conforms to the specifications
- Power circuits are reliably connected
- Liquid is present and running through the pipeline
- Air is not collected in the flowmeter’s location
- Threshold values for the "
Output conversion factor
" and "
Pulse
weight
" as well as for the "
Lower limit by flowrate
" and "
Up. limit by
flowrate
" parameters are set correctly. Change the values, if neces-
sary.
If all the requirements listed above are met contact the service cen-
tre (regional dealer) or manufacturer for the information about the de-
vice’s operability.
4.3.5. In case that the CS, PEA, PS or CS-CP cables are replaced, the flow-
meter must be recalibrated.
28