![Torbal DRX 5 Series Скачать руководство пользователя страница 14](http://html1.mh-extra.com/html/torbal/drx-5-series/drx-5-series_instruction-manual_1130063014.webp)
Instruction Manual - TORBAL DRX 5 Series
7.External Keyboard – This feature must be enabled if an external PC keyboard is connected to the scale via the PS2
port. [Factory Default setting:
Disabled
]
8.Advanced Pill Counting Accuracy (APA) - APA is an advanced pill counting algorithm that can adjust for
production LOT pill weight variations and provide even greater accuracy when counting large quantities. Turn to
page 59 for more information on how to use the APA feature. [Factory Default setting:
Disabled
]
9.Deviation Warning - When enabled the feature allows to check for possible discrepancy between the stored average
piece weight, and the average piece weight of the drug which is counted. The feature signals if a deviation greater
than +/- 2% is found. [Factory Default setting:
Disabled
]
10.Onscreen Instructions - This feature displays instructions that lead you step-by-step through each the pill counting
function. The instructions appear at the bottom of the LCD display. [Factory Default setting:
Enabled
]
11.Default Settings – Selecting this option rests the function configuration and restores factory defaults.
7.2.Compounding – Setup
The compounding setup will allow you to configure the compounding function by enabling or disabling different
components of the function.
1.Access the Main Menu with the MENU key and use Use the up and down arrow keys to select the function to be
configured.
2.Once a desired function is highlighted, press Enter.
14