background image

6000 Series 

 

3.7 Parallel 

Operation 

Parallel connection is used to connect two or more power supplies of the same model  in parallel to obtain a 
higher current output (maximum  24A for this power supply).   

Figure 9 shows the parallel connection between two 6303D power supplies at parallel tracking.  One unit is 
set at 30V/5A (MASTER setting is 30V/2.5A) while the other is at 20V/6A (MASTER setting is 20V/3A).  
Figure 10 shows the parallel connection voltage/current output versus load changes.  

+

-

V

A

+

-

+

-

5V5A

OVER LOAD

I1=IS1+IM1

+

-

V

A

+

-

+

-

5V5A

OVER LOAD

IO = I1 + I2

I2=IS2+IM2

   

SET : V1 = 30V / 5.0A
SET : V2 = 20V / 6.0A

A

3A

9A

12A

10

OHM

8

6

4

2

1

30

10

20

30

V

[V1&V2 C.C. MODE]

[V1 C.C. MODE
 V2 C.V. MODE]

[V1&V2 C.V. MODE]

5A

10

OHM

8

6

4

2

1

30

 

Figure 9   Parallel Connection Diagram 

Figure 10  Parallel Voltage/Current Output vs. Load  

     Change 

NOTE

:  The difference between the voltage settings for two or more power supplies of the same model 

connected in parallel should not exceed 15V.  The minimum voltage setting should not be less 
than 10V.  If it is necessary to be lower than 10V, the voltage difference should be less than 2V.  
The closer to 0V, the smaller the difference required. 

12

 

 

 

Содержание 6302A, 6302AR, 6302D, 6302DR, 6303A, 6303AR, 6303D,6303DR, 6306A, 6306AR, 6306D, 6306DR, 6603A, 6603AR, 6603D, and 6603D

Страница 1: ...6000 Series Operation Manual...

Страница 2: ...pany makes no representations or warranties either expressed or implied with respect to the contents hereof and specifically disclaims any warranties merchantability or fitness for any particular purp...

Страница 3: ...ion 6 3 2 Setting the Output Voltage and Output Current 6 3 3 Constant Voltage Constant Current Characteristics 7 3 4 Operation Modes 7 3 4 1 Independent Mode 7 3 4 2 Serial Tracking Mode 8 3 4 3 Para...

Страница 4: ...serial connection Short circuit protection against external input while providing constant voltage and constant current Allows serial or parallel connection with the same power supply model 5V 5A cons...

Страница 5: ...6000 Series 2 Front and Rear Panels 2 1 Front Panels Front panel with digital display for models 6302D 6302DR 6303D 6303DR 6306D 6306DR 6603D and 6603DR 2...

Страница 6: ...V Mode LED Uses green LED to indicate constant voltage at slave power 2 Power Switch Power ON OFF switch 1 ON 0 OFF 3 C C Mode LED Uses red LED to indicate constant current at slave power 4 SLAVE Cur...

Страница 7: ...utput terminal black 13 Slave GND Terminal Slave ground terminal green 14 Slave Output Terminal Slave positive output terminal red 15 and 18 TRACKING MODE Switches The switch setting label is located...

Страница 8: ...r details 27 D Type 9 pin connector RC1 for Remote Control 28 D Type 9 pin connector RC2 for Remote Control 29 Power Input Socket 30 Input Power Fuse Holder and Input Voltage Selector Acceptable input...

Страница 9: ...t Voltage and Output Current To set the output voltage and current follow these steps 1 Check that the total load to be connected does not exceed the maximum output voltage and current of this power s...

Страница 10: ...e Constant Current Output Voltage Crossover Point Output Current Figure 3 Constant Voltage Constant Current 3 4 Operation Modes 3 4 1 Independent Mode The master and the slave power can be used indepe...

Страница 11: ...POWER switch 3 Turn the SLAVE Current Adjustment Knob clockwise to maximum 4 Set the master output voltage and current as described in section 3 2 The following figures show two different serial oper...

Страница 12: ...e Master Output Terminal or Slave Output Terminal The output current is twice the set master current while the output voltage is the same as the master setting value To operate in the parallel mode 1...

Страница 13: ...utput for use by TTL logic circuits When the load exceeds 5 amperes the red OVERLOAD LED lights up The following figures show the parallel output connection method and its characteristic curve V A 5V5...

Страница 14: ...while the other is at 40V 3A MASTER setting is 20V 3A Figure 8 shows the serial connection voltage current output versus load changes V A 5V5A OVER LOAD V1 VS1 VM1 VO V1 V2 V A 5V5A OVER LOAD V2 VS2 V...

Страница 15: ...load changes V A 5V5A OVER LOAD I1 IS1 IM1 V A 5V5A OVER LOAD IO I1 I2 I2 IS2 IM2 SET V1 30V 5 0A SET V2 20V 6 0A A 3A 9A 12A 10 OHM 8 6 4 2 1 30 10 20 30 V V1 V2 C C MODE V1 C C MODE V2 C V MODE V1...

Страница 16: ...elector unit refer to Figure 2 2 Turn the unit so that the desired input voltage is positioned above the mark 4 Check that the fuse to conform to this new voltage rating Refer to Table 1 below for the...

Страница 17: ...nalog Display A Digital Display D Full scale 3 0 1 2d PROGRAMMING SPEED Rise Time No Load Load 100mS 200mS 6A 500mS 10A 1S 10A Fall Time No Load Load 2 5S 250mS OUTPUT IMPEDANCE 2m 2 H RECOVERY TIME 1...

Страница 18: ...ull scale voltage in Master Current Readback NA 0 to 0 3V for full scale current in Master OPERATION MODE Independent Two independent outputs and 5V fixed output Output from 0 to rating voltage and 0...

Отзывы: